Service Information
4
-
4
10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440
-
466
-
6100 ● FAX 440
-
466
-
6110 ●
a.
If the UUT displayed voltage is not within ±0.001 mV of the Applied
Voltage, press
until the UUT displayed voltage is within
±0.001
mV, adjusting as close to the Applied Voltage as possible.
b.
If the UUT displayed voltage is too far from nominal,
may not
function as expected.
The UUT will typically indicate 8999 or
-999
in this
state. To set the active CAL factor to
its
default value, press the
(1.5s)
arrow buttons simultaneously for 1.5
seconds. Fine
adjustments may still be necessary.
17.
Press UUT
.
The display will change to rES1 [2, 3 …] showing the actual
measured value saved in the previous step.
18.
Press UUT
again. This will increment to the next CAL value.
19.
through
for each remaining value in
below. For
single channel UUTs, omit
Channel 2 values
.
a.
For two channel UUTs using the single
-
ended Copper Mini
-
TC Cable,
disconnect the cable from the UUT Channel 1 input, and reconnect the
cable to the Channel 2 input after completing CAL4
below.
Repeat Step
C
HANNEL
UUT
D
ISPLAY
A
PPLIED
V
OLTAGE
(mV)
1
CAL1
-10
CAL2
(
75
CAL3
-10
CAL4
(
30
2
CAL5
-10
CAL6
(
75
CAL7
-10
CAL8
(
30
Figure 13: Gain Alignment Values
20.
Remove the copper cable from the DC Voltage Source
and UUT.
Cold Junction Compensation Alignment
21.
Connect one
end of the Type E
Thermocouple Cable
to the UUT Channel 1 input.
a.
For two channel
UUTs
using the Split Thermocouple Cable, connect the
miniature thermocouple
connector of one leg to the UUT Channel 1
input, and the miniature thermocouple connector of the other leg
to the
Channel 2 input.
4
If substituting another thermocouple type, use the appropriate Thermocouple Cable for the
selected thermocouple type.