Instrument Description
1-1
10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440
-466-
6100 ● FAX 440
-466-
6110 ●
1.
I
NSTRUMENT
D
ESCRIPTION
Specifications
G
ENERAL
S
PECIFICATIONS
:
Accuracy
±(0.2% R 0.02% Range)
-1
0 to 55 °C
Resistance Ranges and Test Currents:
Full Scale
Resolution
Test Current
10.000 mΩ
1 μΩ
100 mA
100.00 mΩ
10
μΩ
10 mA
1.0000
Ω
100
μΩ
1 mA
10.000 Ω
1 m
Ω
100
μA
100.00 Ω
10 m
Ω
10
μA
Connector Type
M12
Max. Lead Resistance
500 mΩ
per lead
Display
Four and one
-
half
(4 1/2) digit LCD, with Range, Units,
Function, Battery, and Decimal Indicators
Display Backlight
LED Backlight with
30-
second timeout
Reading Rate
3 / Second for Readings, typical
Battery Type
3 AA (IEC LR6, ANSI 15) Alkaline
Battery Life
100
hours standby, 50
hours in 10 mΩ range, typical
Battery Indicator
Four (4)
Stage Battery Charge Indicator
Keypad
Six
(
6
)
m
omentary switches with audible and tactile
feedback
Power Cycle
Configuration
Retention
Instrument retains:
-
Range
-
Test Limit
Maximum Common
Mode Voltage
30 V
Compliance
CE (2014/30/EU) / RoHS (EU 2015/863)
/ EU REACH (EC)
No
.
1907/2006
ESD
IEC 61000
-
4 2:2009, Class B
EMC
EN 55022:2010+A1:2015,
Class A; EN 61000
-4
3:2006+A2:2010, 10 V/m
(80 MHz to 1 GHz)
MIL
-
PRF
-
28800F, Class 2
1
95% confidence interval, approximately 2σ
2
Factors such as range, backlight, and use of the optional display probe will affect battery life