Reference Manual Rev 3.23
Software Versions 3.xx
Page 40
YEAR (Set Year)
•
Range: 2000 to 2099
MONTH (Set Month)
•
Range: 01 to 12
DAY (Set Day)
•
Range: 01 to 31
HOUR (Set Hour)
•
Range: 00 to 23 (24-hour format)
MINUTE (Set Minute)
•
Range: 00 to 59
7.4.8. TEST (Special Test Functions)
Items within this Group allow access to the testing routines for the instrument. With
these routines the scale base output can be monitored and the inputs and outputs
can be tested.
SCALE (Scale Base Test Display)
Verifies the correct load cell capacity and/or load cell wiring is used. It sets up
the instrument as a simple test meter to measure the load cell signal output.
The display reads in millivolts per volt, factory calibrated to 0.1% worst case.
When accessing this item, initially there should be no weight on the scale. In
NSC, OIML or NTEP modes, this display is only active for five seconds before
returning to the menu.
FRC.OUT (Force Outputs)
Forces each of the output drivers in turn. All outputs turn OFF when leaving this
step. The <EDT> key will advance through each output (ie. ON.1 and ON.2).
Pressing <OK> will turn all outputs off and exit the test.
O.LOAD (Overload Count)
Displays a count of the number of times that the A/D conversion has been out of
range since the count was last reset. This value is read only and cannot be
edited.
CLR.OLD (Clear Overload)
Clears the A/D conversion overload counter. The instrument will prompt with
Cont. N. Press <EDT> to change to Cont. Y and <OK> to continue. When
Cont. Y has been chosen the instrument will display DONE to indicate that the
operation has been completed.
7.4.9. FACTRY (Factory Adjustment Menu)
DEFLT (Restore Factory Settings Except for Calibration and Build)
Restores all settings in the digital setup, which are not calibration critical back to
the original new settings installed at the factory. The main use of this routine is
to completely reset an instrument that is being installed on a different scale.