Reference Manual Rev 3.23
Software Versions 3.xx
Page 35
7.4.4. SPEC (Special Settings Menu)
Settings within this group control features including passcodes, key locking, key
functions and display settings.
SAFE.PC (Safe Security Passcode for Digital Setup)
The SAFE.PC (Safe Passcode) allows partial access to Digital Setup (ie. only
non calibration/trade critical settings can be changed). For the Safe Passcode
to have any effect, the FULL.PC passcode must also be set. The default
passcode setting is 000000 that allows free access. Any other number will
enable the passcode functions and restrict access. Refer to Passcodes page
28 and Accessing Setup page 29 for more information.
•
Range 000000 to 999999
•
Default: 000000
FULL.PC (Full Security Passcode for Digital Setup)
The FULL.PC (Full Passcode) can be set to restrict access to Full Digital Setup.
This passcode is used to prevent unauthorised or accidental tampering in the
instrument setup. The default passcode setting is 000000 that allows free
access. Any other number will enable the passcode functions and restrict
access. Refer to Passcodes page 28 and Accessing Setup page 29 for more
information.
•
Range 000000 to 999999
•
Default: 000000
It is important to note that when restricting Full access to Setup the passcode
must not be forgotten. It is only possible to circumvent the passcode at the
factory. Care must be taken when setting the Full Digital Setup Passcode to
ensure that the instrument does not become permanently locked.
KEY.LOC (Front Panel Key Locking)
This item allows individual keys to be locked and unlocked. The display shows
a dash (–) to indicate that a key is locked (inactive) or characters for each key
that is active (ie. the characters P12345 display). The letter P represents the
<POWER> key while the numbers 1234 and 5 represent the remaining
operation keys. The operation keys are numbered from the left with the
<ZERO> key being number 1.
Note: When the <POWER> key is locked, the instrument cannot be turned off
from the front keypad.
•
Default: P12345 - All keys are unlocked (active)
KEY.FN (Key Functions)
The function of the keypad <FUNCTION> key can be selected here. Refer to
Special Functions page 53 for details of the available key functions.
Options are:
NONE
TEST
COUNT
UNITS
HOLD
PEAK.H
LIVE.WT
SHOW.T
HI.RES
A.TARE
SET.PT
•
Default: NONE
AUT.OFF (Auto Power Off / Battery Operation)
The instrument can be set up to automatically power down after a period of no
activity. Weight motion, network communications or any press of the keyboard