45
Accuracy is expressed as ± (percentage of set value + percentage of range)
Function
Referenc
e
Range
Resoluti
on
Accuracy
Remark
DC voltage
100mV
-10.000mV
~
110.000mV
1μV
0.02+0.01
Maximum output current: 0. 5mA
1000mV
-100mV
~
1100mV
10μV
0.02+0.01
Maximum output current: 2mA
10V
-1.0000V
~
11.0000V
0.1mV
0.02+0.01
Maximum output current: 5mA
DC current
20mA
0.000mA
~
22.000mA
1μA
0.02+0.02
External supply for simulate mA:
5V–28V
Maximum load 1KΩ at 20mA
Resistance
400Ω
0.00Ω
~
400.00Ω
0.01Ω
0.02+0.02
Excitation current: ± 0.5–3 mA;
if ± 0.1–0.5, add 0.1Ω;
Accuracy does not include lead
resistance;
4KΩ
0.0000
KΩ
~
4.0000 KΩ
0.1Ω
0.05+0.025
Excitation current: ±0.05
–0.3mA;
Does not include lead resistance;
TECPEL