ENGLISH
21
10
REPORTS - ERRORS
10.2 Errors
In case of an error report, you need to reset the device in order to start a new cycle.
Errors are signaled by the
simultaneous flashing of the DEGAS and SWEEP leds
The type of error is reported on the time LED bar:
All errors are reset by pushing any button for 5"
Display
Causes
Solutions
1M LED
Mains failure during the
cycle
Reset and restart the cycle if
necessary
4M LED
Temperature probe (open)
Contact support to replace the
probe
5M LED
Temperature probe not
working (short circuit)
Contact support to replace the
probe
10.4 General problems
Problems
Possible causes
Solution
Poor cleaning
‐
Presence of non-pretreated
elements (cement, alginates)
‐
Pre-treat with suitable
detergents.
‐
Incorrect cycle settings
‐
Change parameters, increasing
cleaning time.
‐
No noise from ultrasound can
be heard during the cleaning
phase
‐
Please contact technical support
Some instruments
become damaged
‐
Instruments unsuitable for
ultrasonic cleaning (e.g.
mirrors)
‐
Do not sanitize unsuitable
instruments by ultrasonic
cleaning