TEST and CALIBRATION 

PAY  ATTENTION

:  This  procedure  has  to  be  made  with  extreme  attention  and  by  authorized  and  trained  people; 

because starting this procedure it will start both Outputs (relays) causing the activation of connected alarm devices. 
The instrument has three different code protected functions: 

Operation Check, Zero Adjust and Calibration

.  

Operation Check, Zero Adjust, Calibration Check and Calibration

:

 are different code protected functions. To ac-

cess these functions is necessary to insert the relevant “

Code

" through the keys F1s and F2. To have the key pres-

sure recognized, hold pressing it for around a second (

until the Green Led doesn't switch off for a moment

). Then the next key 

can be pressed. In case of error all it takes is waiting around 10 seconds and the sequence is automatically erased. 

Calibration Kit, Sample Gas Bottles

 (

for Calibration Check and Calibration

please, only

 

using a mixture 20%LEL 

(0.88%v/v)

 

Methane in Air (

20.9% Oxygen

).

 

Catalytic sensors cannot work without Oxygen

. It is possible to use 

either  the  disposable  one  litre  cylinders  with  adjust  valve  or  the  high  pressure  ones  with  reduction  gear.  Is  also 
necessary the Tecnocontrol Calibration Kit model TC011. 

“INSTRUMENT OPERATION CHECK” (

Check Code

: F2, F2, F1, F1)

this function allows to effect a functional test 

of the equipment. After having put the system in safety and inserted the "

Code Test

", all Led are switched off and all 

the  relays  are  disarmed.  Then  they  will  switch  on  in  sequence,  the  Led,  from  the  yellow  up  to  the  3°  red.  To  the 
lighting  of  the  various  Led  it  corresponds  the  activation  of  the  relevant  relays  (

relay  "FAULT"  with  the  yellow  Led,  relay 

"ALARM1" with the 1° red Led, relay "ALARM2" with the 2° red Led, relay "ALARM3" with the 3° red Led

). At the end all the Led will 

remain lighted for around 5 seconds, then the central returns at the conditions of normal operation. It is advisable to 
perform this operation every 6-12 months according to the use. 

Note: this function is not working if the 1° and/or the 2° and/or the 3° red Led have already turned on 

 

“ZERO ADJUST” (

Zero Code

: F2, F1, F1, F2)

this function is to adjust the Zero sensor and can be done in clean 

air only (environment without the presence of gas or other pollutants). Immediately after having inserted the "

Zero 

Code

", as a confirmation of the operation carried out there will be 1 I flash of the 1° red Led and the output will be-

come 4,0 mA. We suggest performing this operation  after the installation or after the change of the cartridge and 
every 6-12 months based on the environmental conditions. 

Note

This function is not working if the 2° e/o the 3° red Led have already been switched on .In case the 1° 

red Led is already switched on only instead of flashing it will switch off for confirmation.

 

“CALIBRATION” (

Calibration Code

: F2, F2, F2, F1, F2, F2)

this function allows completely recalibrate the sensor.

 

Warning

: to guarantee that no errors of elaboration happen, the rare possibility exists that during the Calibration 

the 

yellow LED switch off every 8 seconds, 

in this case interrupt the procedure, switch off and witch on the instru-

ment and repeat the Calibration. If condition persists it will be necessary to send the detector to the supplier for the 
reparation. 

Important note:

 

During Calibration routine the mA output indicates 0mA

.

 

The Calibration can be done in clean air only (

environment without the presence of flammable or other polluting gas

). With the 

keys perform the “Calibration

 Code

 ". Wait until the Yellow and Green Led switch on fix and the 1° red Led starts to 

flash. Insert the TC011 into the sensor holder to regulate the influx of the gas the way that the flow meter indicates 
around 0.3 l/mins (

see Fig.3

). Wait around 3 minutes, then when the 3° red Led switch on (

and while it's SWITCHED ON

), 

press the key F2 on the instrument and hold it pressed until 1° and 3° Red Led are switched off for at least 2 sec-
onds (

if the 1° Red Led continue to flash, wait that the 3° Red Led switch on again and repeat the operation

). 

Then, close the gas cyl-

inder and remove TC011

. At this point we can have two possibilities: 

Yellow  and  Green  Led  illuminates:

  the  calibration  routine  has  correctly  been  performed. 

Wait  8  seconds,  until 

the instrument automatically restores the normal working conditions

(

see “Operational Description > Preheating”

Yellow LED illuminates:

 the routine has failed. In this case, wait 8 seconds, until the instrument automatically re-

peat 

Preheating

, then repeat the “

Calibration”

 routine without inserting again the code. If condition still persists after the 

replacement of the cartridge, it will be necessary to send the detector back to the manufacturer for reparation. 

“CALIBRATION  CHECK”  (

Cal Check Code

:  F2,  F1,  F2,  F1)

:

 

this  operation  allow  to  effect  a  real  functional  test  of 

the equipment with gas

 

after the "

Calibration

" routine, or after the installation. The “

Calibration Check

” routine should 

be done during the periodic maintenances because this is the only method to verify the instrument real functioning. 
With the keys perform the "

 Cal Check Code

 ". Wait until the Yellow Led starts to flash (the green one remain fix) In-

sert the TC011 over the sensor holder, adjust the sample gas bottle valve as the flow meter indicates around 0.3 
l/mins (

see Fig.3

), wait for 3 minutes and check that the instrument activates Alarms as shown in chapter “

Operational 

Description

”. 

If the result is different, is necessary to recalibrate the sensor (

see “Calibration”

).

 Then, close the gas bot-

tle, remove TC011, 

press the key F2 on the instrument to restore the normal working conditions

NOTE

:  further  to  the  above,  If  the  mA  output  is  also  used,  verify  with  a  voltmeter  connected  to  the  Test-Point 

TESTmA

”, the value reach a value between 

184 and 216mV

, [corresponding to 20mA (±1.6) output and the central 

unit should be display about 20% LEL (±2)]. Then, the mA output will slowly decrease up to 4mA. 

 

FR

 

 DESCRIPTION 

Le SE137K est un détecteur ponctuel pour gaz et vapeurs combustibles avec capteur catalytique pouvant égale-

ment être utilisé en systèmes centralisés d'alarme pour l’industrie et le tertiaire. Le détecteur est constitué par une 
boîtier contenant le circuit électronique et les borniers de raccordement; dans le porte capteur, placé dans la partie 
inférieure du boîtier, est  logée  une "

cartouche capteur échangeable

" contenant l'élément sensible et les  données 

identificatrices et de réglage. Le SE137K possède 3 relais d'alarmes réglables en % de la LIE (

Limite Inférieur d'explo-

sivité

), intervenant comme indiqués au 

Tableau 2

 et 1 relais de dérangement. 

Combustion 

Control 

ltd

Summary of Contents for SE137K Series

Page 1: ...e relè relays relais 24V 1A SPST 4 20 mA lineare Linear linéaire Campo di misura Standard Range Champ de mesure 0 20 LIE LEL Limite Scala Limits Limite échelle 30 LIE LEL Vita media in aria pulita Average Life in fresh air Vie moyenne en air pur 5 anni years ans Tempo di risposta T90 Response Time T90 Temps de réponse T90 60 secondi seconds secondes Ripetibilità Repeatability Répétitivité 5 del se...

Page 2: ...e il tasto F2 sulla targa per ripristina re le condizioni di funzionamento normale NOTA In aggiunta a quanto sopra se si usa anche l uscita in mA controllare anche con i puntali del volmetro sui Test Point TESTmA vedi Fig 2 si raggiunga un valore tra 184 e 216 mV ovvero che l uscita in mA aumenti fino a circa 20 mA 1 6 e la centrale cui è collegato il Trasmettitore indichi circa 20 LIE 2 Poi l usc...

Page 3: ...f electric devices in areas with danger of explosion Mounting The Fig 2 shows the instrument size The unit must be positioned vertically with the sensor downwards Models SE137KG and SE137KB should be fixed at 20 30 cm from the floor LPG and Petrol vapours are heavier than air Models SE137KM and SE137KI should be fixed at 20 30 cm from the ceiling Methane and Hydrogen are lighter than air Electrica...

Page 4: ...t and repeat the Calibration If condition persists it will be necessary to send the detector to the supplier for the reparation Important note During Calibration routine the mA output indicates 0mA The Calibration can be done in clean air only environment without the presence of flammable or other polluting gas With the keys perform the Calibration Code Wait until the Yellow and Green Led switch o...