Wide-range density specification (special density calibration)
If the process temperature is outside the valid range (→ 261) the measured error is
±0.00005 g/cm
3
/°C (±0.000025 g/cm
3
/°F)
[kg/m ]
3
[°C]
0
- 0
8
80 160 240
[° ]
F
- 0
5
0
100
200
300
0
2
4
6
8
10
2
1
12
14
16
18
320 400 480 560 640
A0027453
1
Field density calibration, for example at +20 °C (+68 °F)
2
Special density calibration
Temperature
±0.005 · T °C (± 0.005 · (T – 32) °F)
Influence of medium
pressure
The table below shows the effect on accuracy of mass flow due to a difference between
calibration pressure and process pressure.
o.r. = of reading
It is possible to compensate for the effect by:
• Reading in the current pressure measured value via the current input.
• Specifying a fixed value for the pressure in the device parameters.
Operating Instructions .
DN
[% o.r./bar]
[% o.r./psi]
[mm]
[in]
8
3/8
no influence
15
½
no influence
25
1
no influence
40
1½
–0.003
–0.0002
50
2
–0.008
–0.0006
80
3
–0.009
–0.0006
100
4
–0.007
–0.0005
150
6
–0.009
–0.0006
250
10
–0.009
–0.0006