CARDIATEK UG 202001-R01
Maintenance Guide
CLEANING THE CARDIATEK SYSTEM (CONTINUED)
CLEANING SOLUTIONS CONTINUED
WARNING
9
Some cleaning products are corrosive in nature and may cause damage to the product if used improperly.
9
If the products described are used to clean the equipment, the Cardiatek System and components must be rinsed
with clean water and thoroughly dried following cleaning. Failure to properly rinse and dry the system will leave
a corrosive residue on the surface of the product, possibly causing premature corrosion of critical components.
REMOVAL OF IODINE COMPOUNDS
Use a solution of 1/4 Tablespoon of Sodium Thiosulfate in 1 pint / .5 litre of warm water to clean the stained area.
Clean as soon as possible after staining occurs. If stains are not immediately removed, allow solution to soak or stand
on the surface. Thoroughly rinse surfaces with clean water and allow to air dry before returning unit to service.
WARNING
9
Failure to properly clean contaminated parts or components will increase the risk of exposure to blood borne
pathogens and may cause injury to the patient or the operator.
PREVENTIVE MAINTENANCE
A preventive and regular maintenance program should be established for all Technimount equipment. Preventive
maintenance may be required more frequently based on the usage level of the product. The Cardiatek System requires
regular preventative maintenance. Establish and follow a preventive maintenance schedule and keep records of
preventive maintenance activities (see maintenance form in this user guide).
CAUTION
9
Close attention should be given to safety features including, but not limited to the handle and locking mechanisms
and Discs.
9
Improper maintenance can cause injury or damage to the product. Maintain the product as described in this user
guide.
LUBRICATION
The Cardiatek System has been designed to operate without the need for lubrication. Due to an anodization of the
metallic parts, using a silicone-base lubricant can be used but it is not necessary. If you are not sure what to use, please
contact Technical Support at
.
CAUTION
9
Failure to use authorized parts, lubricants, etc. could cause damage to the cot/stretcher and will void the warranty
of the product.
Summary of Contents for Cardiatek
Page 1: ...CARDIATEK USER GUIDE ...
Page 43: ...www technimount com CARDIATEK UG 202001 R01 43 Return To Table of Contents SAE Certification ...
Page 44: ...Return To Table of Contents 44 CARDIATEK UG 202001 R01 www technimount com FAA Certification ...
Page 45: ...www technimount com CARDIATEK UG 202001 R01 45 Return To Table of Contents CE Certification ...