background image

www.technimount.com

BP151 UG 202108-01

21

Return to table of content

MAINTENANCE GUIDE

CLEANING (CONTINUED)

CLEANING SOLUTIONS

Phenolic type or quaternary type disinfectants can be used (excluding Virex® TB). Iodophor type disinfectants, Peracetic 

Acid, Chlorine, and especially Hydrogen Peroxide are not recommended for use because staining and corrosion may 

result.

Suggested cleaners for the bracket: 

 

9

Quaternary Cleaners (active ingredient - ammonium chloride)

 

9

Water and soft soap

 

9

A mild 10% bleach solution, a solution containing alcohol or a quaternary ammonium chloride solution. In diluted 

concentrations, these cleaning agents have no negative effect on our products.

WARNING

 

9

Some cleaning products are corrosive in nature and may cause damage to the product if used improperly.

 

9

If the products described are used to clean the equipment, the bracket and components must be rinsed with clean 

water and thoroughly dried following cleaning. Failure to properly rinse and dry the system will leave a corrosive 

residue on the surface of the product, possibly causing premature corrosion of critical components.

REMOVAL OF IODINE COMPOUNDS

If rust appears on any surface of the product, we recommend sanding the affected surface to remove the rust and 

then using the cleaning method presented in this guide once the rust is removed. To clean the stained area, use a 

solution of 1/4 Tablespoon of Sodium Thiosulfate in 1 pint / .5 litre of warm water to clean the stained area. Clean as 

soon as possible after staining occurs. If stains are not immediately removed, allow solution to soak or stand on the 

surface. Thoroughly rinse surfaces with clean water and allow to air dry before returning unit to service.

WARNING

 

9

Failure to properly clean contaminated parts or components will increase the risk of exposure to blood borne 

pathogens and may cause injury to the patient or the operator.

PREVENTIVE MAINTENANCE

A preventive and regular maintenance program should be established for all Technimount equipment. Preventive 

maintenance may be required more frequently based on the usage level of the product. The bracket requires regular 

preventative maintenance. Establish and follow a preventive maintenance schedule and keep records of preventive 

maintenance activities (see maintenance form in this user guide). 

CAUTION

 

9

Close attention should be given to safety features including, but not limited to locking mechanisms and discs.

 

9

Improper maintenance can cause injury or damage to the product. Maintain the product as described in this user 

guide.

Summary of Contents for BRACKET PRO 151 Series

Page 1: ...BRACKET PRO SERIE 151 USER GUIDE ...

Page 2: ...p with Maxtec Maxtec does not promote or sell Technimount products In order to ensure continuous improvements to our products and services we invite you to communicate your comments to our Customer Service Department at customerservice technimount com LEGAL NOTICE Technologies CGC Inc or Technimount E M S Holding Inc or its affiliates are not expressly associated with the marketing distribution or...

Page 3: ...our Technimount product its components or for any technical questions during the installation or maintenance processes please send an email to techsupport technimount com Please have the serial number of your Technimount product available as shown in the figure below when calling Technimount Technical Support Include the serial number in all written communications SERIAL NUMBER LOCATION CONTACT IN...

Page 4: ...NOTICE CERTIFICATION 14 INSTALLATION GUIDE 15 Removing the packaging for installation 15 Tools Required for Installation 15 Identification of Components 16 U Block Support 16 hardware kit 16 Installation of the Bracket onto the MaxBlend2 Oxygen Blender 17 MAINTENANCE GUIDE 20 Cleaning 20 Cleaning Process 20 Removal of Iodine Compounds 21 Preventive Maintenance 21 Inspection Process and Schedule 22...

Page 5: ...mount com BP151 UG 202108 01 5 TABLE DES MATIÈRES Return Authorization 27 RETURN POLICY 27 Damaged Merchandise 28 Claim Process 28 Required Information 28 Findings and Conclusion 28 Questions about our Policy 28 ...

Page 6: ... 6 BP151 UG 202108 01 www technimount com INTRODUCTION SYMBOLS AND DEFINITIONS SYMBOL DESCRIPTION Warning and Caution special attention is required Consult accompanying documents Safe working load symbol and Load balance symbol ...

Page 7: ...serious injury It may also describe potential serious adverse reactions and safety hazards Caution Alerts the reader of a potentially hazardous situation which if not avoided may result in minor or moderate injury to the user or patient or damage to the equipment or other property This includes special care necessary for the safe and effective use of the device to avoid damage that may occur as a ...

Page 8: ...ww technimount com PRODUCT ILLUSTRATION BRACKET PRO SERIE 151 Front View INTRODUCTION Top Support Block to secure Medical Device Support U Block for Installation at the Back of Medical Device High Density Aluminum Structure with Black Anodized Finish ...

Page 9: ... to table of content PRODUCT ILLUSTRATION CONTINUED BRACKET PRO SERIE 151 CONTINUED Back View INTRODUCTION Rubber Cushions Protecting the Medical Device into the Bracket Anti Rotation Micro Disc for Installation on Vertical and Wall Mounting Systems ...

Page 10: ... and install medical equipment owned by the customer on an incubator transport stretcher manufactured by Stryker and involving or not a Dräger or other incubator brand The purchase and use of products designed by Technimount are subject to this limitation of liability clause It is important to understand that as Technimount is not the manufacturer of the Stryker stretcher or other brand nor the ma...

Page 11: ... Device Retained on Main Assembly with Support Blocks Attachment Options Anti Rotation Micro Disc Transportation Compatibility System Compatible with Technimount s Safety MD Transporter vertical Vertical Mounting System Micro Base Features Access to all Medical Device Monitors Connectors and Accessories Rubber Cushions Protecting the Medical Device into the Bracket Anti Rotation Micro Disc for Ins...

Page 12: ... fall during transport and may cause injury to the patient or operator and or damage to the product 9 Do not allow untrained staff to assist in the operation of the mounting system Untrained technicians staff can cause injury to the patient or themselves 9 Do not place additional equipment or items onto the bracket or mounts other than approved Technimount or components as this could damage the pr...

Page 13: ...ossibly causing premature corrosion of critical components 9 When ordering it is important to identify on which mounting system you need to mount the bracket on as there are different systems for different applications ground 9 A maintenance program should be established for all Technimount equipment Preventive maintenance may need to be performed more frequently based on the level of use of the p...

Page 14: ...ount com WARNING The end user and administrator are responsible to validate regulations and standards for safety in their region in order to comply with applicable safety regulations Technimount is not responsible to inform the end user and or administrator of applicable legislation for safety in their area LEGAL NOTICE The purchase and use of products designed by Technimount are subject to this l...

Page 15: ...if needed TOOLS REQUIRED FOR INSTALLATION Below are the basic tools required for installation Other tools may also be required If you have any questions regarding the installation or need assistance please contact our Technical Support team at techsupport technimount com 9 Phillips Screwdriver NOTE 9 Electrical tools are not recommended as there is a potential risk of damage to the threads 9 Techn...

Page 16: ...PPORT The black or silver Stainless Steel support U block is used for installation on the rear block of the MaxBlend2 Oxygen Blender It secures the medical device HARDWARE KIT Thefollowinghardwareisprovidedfortheinstallation of the Bracket Pro Serie 151 9 Four 4 Black Oxide Stainless Steel Flat Head Phillips Screws 10 32 x 1 2 ...

Page 17: ...lled on the Maxtec MaxBlend2 Oxygen Blender 2 Locate the rear block of the medical device 3 Unplug the bottle before installing the main structure 4 Locate the U Block of the main assembly and align with the rear block of the medical device INSTALLATION GUIDE INSTALLATION OF THE BRACKET ONTO THE MAXBLEND2 OXYGEN BLENDER ...

Page 18: ...ce and the U block already installed 7 Install the main assembly at the rear of the medical device 8 Align the screws and screw holes from the main assembly 9 Install the Four 4 Black Oxide Stainless Steel Flat Head Phillips Screws 10 32 x 1 2 with the U block tightening will be at the end NOTE Put a few drops of medium strength thread locking adhesive blue Loctite 242 on each screw before install...

Page 19: ...O THE MAXBLEND2 OXYGEN BLENDER CONTINUED 10 Tighten the Four 4 Screws 11 Plug the bottle back CAUTION 9 Do not use electrical tools to install the screws 9 Do not overtighten the screws 12 The Maxtec MaxBlend2 Oxygen Blender is ready for transport once properly inserted into its bracket and the stretcher mounting system ...

Page 20: ...scoloration from continuous washing however degradation of the product s performance or functionality will not be affected due to power washing if cleaning instructions are properly followed The material used to manufacture the bracket is high grade aluminum it is very durable and resistant to corrosion however high levels of sodium or a high concentration of sodium or acid may prematurely corrode...

Page 21: ... the product we recommend sanding the affected surface to remove the rust and then using the cleaning method presented in this guide once the rust is removed To clean the stained area use a solution of 1 4 Tablespoon of Sodium Thiosulfate in 1 pint 5 litre of warm water to clean the stained area Clean as soon as possible after staining occurs If stains are not immediately removed allow solution to...

Page 22: ...erform these inspections please contact your Technical Support Team at techsupport technimount com Check each routine and replace damaged or worn out parts if necessary Follow timetable below for maintenance either in months M or hours H in accordance with your Service s current maintenance practices and protocols ITEM ROUTINE 1 M OR 2 H 3 M OR 6 H 6 M OR 12 H 12 M OR 24 H Bracket 9 Inspect the br...

Page 23: ...www technimount com BP151 UG 202108 01 23 Return to table of content GUIDE D ENTRETIEN INSPECTION AND MAINTENANCE RECORD PREVENTIVE MAINTENANCE PERFORMED BY DATE TIME ...

Page 24: ...Return to table of content 24 BP151 UG 202108 01 www technimount com MAINTENANCE GUIDE TRAINING RECORD DATE TRAINING METHOD TRAINEE NAME BASIC TRAINING TRAINING UPDATE ...

Page 25: ... identified on the assembly drawing parts in this user guide may not be individually available for purchase Please call Technimount Customer Service 1 888 639 2758 or at customerservice technimount com for availability and pricing DESCRIPTION PART NUMBER Bracket Pro Serie 151 2010 10 MXBL Four 4 Black Oxide Stainless Steel Flat Head Phillips Screws 10 32 x 1 2 TCM VF 999 038 U Block TCM 102 003 ...

Page 26: ...aced after the Warranty Period has expired the Warranty Period for the repair or replacement will expire three 3 months after the date of repair or replacement LIMITED RESPONSIBILITY AND WARRANTY The Limited Warranty does not apply to normal wear that should result from normal use It does not apply when the Hardware or any component is opened disassembled or repaired by someone not authorized by T...

Page 27: ... 45 DAYS Product must be unused undamaged and in its original packaging The product return request must be provided in writing and it must be approved by TECHNIMOUNT prior to returning the product Returns will not be approved on modified or damaged item Charges may apply if damaged or missing items Customer is responsible for a 25 restocking fee PRIOR 60 DAYS Product must be unused undamaged and i...

Page 28: ...the RMA form online at https www technimount com service support service support overview REQUIRED INFORMATION Serial number Invoice number Identification of the issue Upload the pictures of the issue Submit the form A confirmation of the receipt of the claim will be sent to you within 2 business days If you have not received a confirmation email by then please contact techsupport technimount com ...

Page 29: ......

Page 30: ...t matters most Technimount E M S Holding Inc 3770 Jean Marchand Street Suite 100 C Quebec QC G2C 1Y6 Canada T 1 888 639 2758 F 1 855 339 6351 technimount com info technimount com 2021 Technimount E M S Holding Inc All Rights Reserved Printed in Canada ...

Reviews: