Troubleshooting 

1

Refer to LT– 38213, Inline Cryogenic Filter Instructions, for cleaning or replacement. 

 
 

If the problem continues contact Technifab Products, Inc for further assistance. 

Problem Action 

System will not power up or 
screen is dark 



Check electrical supply to make sure there is 
power to the unit. 



Unplug/ reset the Control Box by unplugging 
the unit.  

Unit switches to next tank before 
active tank is empty.  



Check to make sure all cables to inlet valves 
are connected. A small LED in the solenoid 
connector indicates the valve should be on. 



If the pipe system is at room temperature, It 
takes longer to deliver liquid nitrogen. Press 
the onscreen Run/Stop button, press the 
onscreen reset button, and press the Run/Stop 
button again. A longer Switcher Timer SP 
setting may be necessary. 

Error message on screen 



Check all wires and connectors. 



On Initial Setup an error may display 
“RGEF130: Input Data error”. This is caused 
either by wires not attached correctly or the 
value is too low to read by the PLC. Press “X” 
and continue to monitor system closely. 

Slow fill 



Check supply dewars for liquid and pressure. 



Check the screen to make sure the “Active” 
light is on for a tank that has liquid nitrogen 
and the tank valve is open. 



Make sure the tank switcher inlet filter (if 
equipped) is not plugged or clogged. 

1

 

No fill 



Check to make sure the active tank has liquid 
nitrogen and is turned on.  



Make sure the tank switcher inlet filter (if 
equipped) is not plugged.

1

 

Summary of Contents for Cryogenic Tank Switcher

Page 1: ...LT 33020 2 2013 Technifab Products Inc 1 Technifab Vacuum Jacketed Tank Switcher Installation Manual...

Page 2: ...013 Symbols Used in this Manual Warning indicates a potentially hazardous situation which if not avoided could result in death or serious injury Caution indicates a potentially hazardous situation whi...

Page 3: ...chnifab Products Inc recommends routing the discharge lines from pressure relief devices to a safe place of discharge The outlet of the discharge line should not be in a location where it could injure...

Page 4: ...her Manifold Assembly with Junction Box Note Most tank switchers include a cryogenic inlet filter installed inside the inlet female bayonets You should use an orange colored bayonet clamp two shown to...

Page 5: ...ld be large enough to support the manifold The photo shows an example of a well mounted tank assembly Be sure to mount the manifold in three places shown with arrows on in the photo one clamp on each...

Page 6: ...cables are connected on the right side from the bottom up 6 Hookup the inlet and outlet piping to the manifold If the tank switcher inlets have inline cryogenic filters be sure to use the supplied ora...

Page 7: ...econfigured at the factory For shop air the Pneumatic Actuated Valve Guidelines Technifab part number literature LT 35569 has a more complete description of the differences and shop air requirements F...

Page 8: ...deliver liquid nitrogen Press the onscreen Run Stop button press the onscreen reset button and press the Run Stop button again A longer Switcher Timer SP setting may be necessary Error message on scre...

Page 9: ...tts 110 VAC nominal 1 7 amp maximum Alarm Output 24 VDC 0 2 amp max current Product Specifications Technifab Cryogenic Tank Switcher Support Information Technifab continuously seeks to improve its pro...

Reviews: