PAGE 10
T E C A L 1 4 0 H O P E R A T O R ’ S M A N U A L
Specification
The figures stated in the specification are as measured 0-50 mm from the
base of the well of the insert, using an SPRT, at the time of calibration.
Measuring zone 0 to 50 mm
0" to 2"
from base of well
from base of well
Depth of well in insert
114.3 mm
4.5"
Temperature range
-40°C to 140°C *
-40°F to 284°F *
Temperature accuracy in measuring zone
±0.3°C
±0.54°F
Temperature uniformity in measuring zone ±0.2°C at 100°C
0.36°F at 182°C
Temperature stability after 10 minutes
±0.05°C
±0.09°F
Display resolution
0.1°C
0.1°F
Heat up rate
5 minutes
20°C to 100°C
68°F to 182°F
Cool down rate
9 minutes
100°C to 0°C
182°F to 32°F
Programmable ramp rate
0.1 to 10°C/min
0.2 to 18 °F/min
Switch Test
Yes
Comms Port 9way D type
bi-directional RS232
* The lowest temperatures can only be achieved when the Tecal 140 is used with
a chilled water supply which has a maximum pressure of 6 metres of water.
However, the chilled water supply must not be switched on when the set
temperature is above 40°C as this can lead to damage to the block.
Dimensions
Height
272mm
10.7"
Width
185mm
7.28"
Depth
439mm
17.28"
Wells in the block:
Central Well for inserts, see the list of inserts at the back of this book
Depth
130mm
5.1"
Diameter
38.2mm
1.5"
Working conditions
The Tecal unit is designed to work safely under the following conditions:
Ambient temperature range
5°C to 40°C
Humidity
Up to 95% relative humidity, non-condensing
Note:
The control specifications quoted are for an ambient temperature
range of 10°C to 30°C. The specification may deteriorate outside this range
but the unit will still work safely.
Radio frequency interference tested and passed to EN50081-1.
Immunity Tested and passed to EN50082-1