For all models, bath temperature is monitored and controlled by a PRT in conjunction with a 3 term controller.
Protection in all the units is provided by means of an adjustable over-temperature cut-out. The pump motor and the 

transformer are also fitted with thermal fuses.

Techne Baths

Techne manufacture a range of unheated baths from 8 to 48 litre capacity which can be fitted with any of the 

thermoregulators. The B-8, B-12, B-18, and B-26 bath inner containers are manufactured from stainless steel for 
maximum corrosion resistance and are deep drawn with large easy clean corner radii. The B-48 inner is also manu-
factured from stainless steel but is of welded construction. These baths are suitable for use up to 200°C. 

For higher temperatures (up to a maximum of 250°C) Techne has a range of liquid calibration baths of capacities 

5, 7 or 12 litres. These baths are fully insulated providing excellent temperature retention and stability and are ideal 
for calibration purposes.

In addition, Techne offer a range of refrigerated baths of 7, 12 or 22 litre capacity. These have a built-in refrigeration 

system for temperatures down to -35°C.

For further information, please contact [email protected].

Safety and Installation

Introduction

Please read all the information in this booklet before using the unit.

 Safety  Advice

Techne have taken great care in the design of these units to protect operators from hazards, but operators should 

pay attention to the following points:

v  

HIGH TEMPERATURES ARE DANGEROUS: they can cause serious burns to operators and ignite combustible material.

USE CARE AND WEAR PROTECTIVE GLOVES TO PROTECT HANDS.

DO NOT use combustible substances near hot objects.

DO NOT operate the unit in the vicinity of flammable liquids or gases.

NEVER lift or carry the unit until it has been switched off and allowed to cool for at least 30 minutes.

The unit should be carried using both hands.

NEVER move or carry the unit when in use or connected to the mains electricity supply.

DO NOT position the unit so that it is difficult to disconnect from the mains supply using the mains plug.

The mains outlet socket used should be located close to the equipment and readily identifiable and

accessible to users.

Operator Safety

All operators of Techne equipment must have available the relevant literature needed to ensure their safety. It is im-

portant that only suitably trained personnel operate this equipment, in accordance with the instructions contained 

in this manual and with general safety standards and procedures. If the equipment is used in a manner not specified 

by Techne the protection provided by the equipment to the operator may be impaired.

All Techne units have been designed to conform to international safety requirements and are fitted with an

over-temperature cut-out. On some models, the cut-out is adjustable and should be set to suit the application. On 

all other models the cut-out is pre-set to protect the unit.

If a safety problem should be encountered, switch off at the mains socket and remove the plug from the supply.

Summary of Contents for 083010

Page 1: ...0 Heating element with thermostat and stirrer TE 10D Nijv Nijverheidsstr erheidsstraat 9 aat 9 698 6987 EN Giesbeek The Netherlands 7 EN Giesbeek The Netherlands T T 31 313 880 200 31 313 880 200 E E...

Page 2: ...moregulator 24 The Front Panel Controls 25 After Use 27 Setting the Over temperature Cut Out 28 RS232 Serial Interface 29 Calibration of the Thermoregulator 30 Service and repair 30 Warranty 30 Techni...

Page 3: ...recommended by the manufacturer the protection provided may be impaired General Description The Techne thermoregulators are designed to be used with any Techne unheated bath liquid calibration bath o...

Page 4: ...ay attention to the following points v HIGH TEMPERATURES ARE DANGEROUS they can cause serious burns to operators and ignite combustible material v USE CARE AND WEAR PROTECTIVE GLOVES TO PROTECT HANDS...

Page 5: ...unit 4 Do not switch on until the unit is fully installed see page 19 5 Note that the following symbols may be next to the indicator lamps on the front panel of the units and have the following meani...

Page 6: ...Decontamination Certificate supplied with your unit Cleaning v Before cleaning your unit ALWAYS disconnect it from the power supply and allow it to cool below 50 C v Your unit can be cleaned by wiping...

Page 7: ...l op rateur Tous les utilisateurs de produits Techne doivent avoir pris connaissance des manuels et instructions n cessaires la garantie de leur s curit Important cet appareil doit imp rativement tre...

Page 8: ...pt de d fauts de mat riaux et de fabrication lorsqu il est utilis dans des conditions normales de laboratoire pendant la p riode sp cifi e sur la carte de garantie ci jointe En cas de r cla mation jus...

Page 9: ...euse pour nettoyer l appareil Veillez ne pas introduire d eau dans l appareil N utilisez pas de produits abrasifs Coupe circuit d exc s de temp rature v En l absence de puissance de chauffe v rifiez l...

Page 10: ...gang zu der entsprechenden Literatur haben um ihre Sicherheit zu gew hren Es ist wichtig da diese Ger te nur von entsprechend geschultem Personal betrieben werden das die in dies er Gebrauchsanweisung...

Page 11: ...s defekte Teil oder das ganze Ger t kostenfrei ersetzen Die Garantie VERF LLT bei v Jeglichem Reparaturversuch durch jemand anderen als den Hersteller oder seinen Vertreter v Auftreten jeglicher klein...

Page 12: ...zen Sie dann den bertemperaturabschalter an der R ckseite des Ger tes wieder zur ck indem Sie den roten Knopf einmal bis zum Anschlag dr cken v Wenn der bertemperaturabschalter wiederholt ausl st lieg...

Page 13: ...nual y con las normas y procedimientos generals de seguri dad es muy importante que s lo personal debidamente capacitado opere estos aparatos De no ser as la protec ci n que el equipo le proporciona a...

Page 14: ...e alguno Esta garant a NO se aplicar si v Se ha realizado o intentado cualquier reparaci n por otra persona distinta al fabricante o sus agentes v Aparece cualquier ara azo o esquirla menor en el recu...

Page 15: ...Si el calefactor no recibiera alimentaci n compruebe el enchufe y el cable de la toma de corriente a continuaci n vuelva a ajustar el control del dispositivo si su equipo lo lleva montado v Una desco...

Page 16: ...ra Techne deve avere a disposizione la documentazione necessaria al fine di assicurare la loro incolumit importante che solo personale adeguatamente addestrato utilizzi questo apparecchio in conformit...

Page 17: ...llegata In caso di reclamo ragionevole TECHNE sostituir qualsiasi componente difettoso o sostituir l unit senza spese aggiuntive La presente garanzia NON si applica in caso di v Riparazioni effettuate...

Page 18: ...l apparecchio controllare la spina elettrica e il relativo cavo collegati alla rete v Ripetute interruzioni del funzionamento dell apparecchio indicano un serio malfunzionamento in questo caso restitu...

Page 19: ...ally Screw the pipe connectors supplied with the unit to connect the appropriate pipe to the unit Set for internal circulation Set for external circulation 2 CAUTION DO NOT SWITCH THE THERMOREGULATOR...

Page 20: ...he corners of the bath 6 Locate the thermoregulator above the bridge piece Toggle the heating element carefully through the clearance hole Lower the thermoregulator onto the M4 threaded stud 1 and 3mm...

Page 21: ...s on page 36 The cooling coil is fitted to the thermoregulator just below the over temperature cut out switch You will find a rubber moulding screwed to the base and filling two slots Remove the screw...

Page 22: ...de ionised with neutral pH 10 C to 150 C Dow Corning Silicone Oil 200 series 10 C to 250 C Dow Corning Silicone Oil 210H 100cs series Warning check gel life at top end of range Extraction may be neces...

Page 23: ...oses should be fitted to the outlet and return pipes A suitable hose must be capable of withstanding both the temperature of operation and the liquid being used Always securely clip the hoses in place...

Page 24: ...e POWER indicator on the front will light up When Switching On The Thermoregulator When first switched on the display will show the edition of the software currently installed For example software iss...

Page 25: ...the SET temperature button is held down and the UP ARROW button is pressed the set temperature is in creased The DOWN ARROW button When the SET temperature button is held down and the DOWN ARROW butt...

Page 26: ...erature indicator As the temperature approaches the set temperature the heater indicator will flash When the set temperature is reached the indicator will stay on for shorter periods If the bath tempe...

Page 27: ...gin to return to ambient even if the heater light is on the light staying on or not depends on which circuit has sensed a fault After Use When you have finished heating samples remember that parts of...

Page 28: ...Return the unit to the liquid switch on the mains power The heater will work again Or Allow the liquid to cool may be as much as 40 C the heater will cool and reset you will hear an audible click The...

Page 29: ...h the unit and the PC before either unit is powered up otherwise data integrity cannot be guaranteed Once the cable is fitted it does not matter which unit is powered up first The following tables ind...

Page 30: ...same temperature as the measured value 4 Press UP and DOWN together to confirm the value The display will return to the bath temperature and the unit will control with the new calibration parameters S...

Page 31: ...et point accuracy 1 C 1 C 1 C Method of control PID PID PID Temperature sensor PRT PRT PRT Nominal heater power 230V 1000W 1800W 1800W 120V 1000W 1500W 1500W Maximum watts density 7 6 W cm 6 2 W cm 6...

Page 32: ...ll height C 172 mm 172 mm 172 mm 222 mm 298 mm Internal Maximum length D 240 mm 329 mm 505 mm 505 mm 560 mm Working length E 115 mm 205 mm 380 mm 380 mm 430 mm Width F 300 mm 300 mm 300 mm 300 mm 330...

Page 33: ...ll Width B 260 mm 260 mm 260 mm Overall Height C 183 mm 233 mm 358 mm Opening length D 140 mm 140 mm 140 mm Opening width E 140 mm 140 mm 140 mm Maximum working depth F 130 mm 140 mm 305 mm Minimum wo...

Page 34: ...width B 250 mm 370 mm 395 mm Overall height C 566 mm 610 mm 565 mm Internal Maximum length D 192 mm 208 mm 360 mm Working length E 224 mm 224 mm 250 mm Width F 151 mm 300 mm 295 mm Maximum working de...

Page 35: ...number Description 6007349 O ring thermometer 6103913 Cap seal 6103475 Cap 6103804 Pipe connection seal 6103771 Pipe connection nozzle 6103460 Bottom outlet nozzle I FCLAMP Clamp for TE 10D I FCLAMP2...

Page 36: ...BLE48 Adjustable Trays Supported by a ball chain and clip the stainless steel trays can be used to alter the depth of the unheated baths Bath Size Part Code 8 litre FADJ08 12 litre FADJ12 18 litre FAD...

Page 37: ...table for use with the TU 20HT models Product Code Description FFC2D FC 200 flow cooler 20 C 230V FFC2P FC 200 flow cooler 20 C 115V FFC5D FC 500 flow cooler 35 C 230V FFC5P FC 500 flow cooler 35 C 11...

Page 38: ...38...

Page 39: ...tive Beacon Road Stone Staffordshire ST15 0SA United Kingdom Object of Declaration Thermoregulator reference the attached list of catalogue numbers The object of the declaration described above is in...

Reviews: