Techmed TMA-6000 Operation Manual Download Page 13

Template version:1 

13/15

Table 2 

For all ME EQUIPMENT and ME SYSTEMS 

Guidance and manufacturer’s declaration - electromagnetic immunity 

The TMA-6000 is intended for use in the electromagnetic environment specified below. The 
customer or the user of the TMA-6000 should assure that it is used in such an environment.

IMMUNITY test 

IEC 60601test level 

Compliance level 

Electromagnetic 

environment - 

guidance 

Electrostatic 
discharge (ESD) 
IEC 61000-4-2 

± 6 kV contact 
± 8 kV air 

± 6 kV contact 
± 8 kV air 

Floors should be 
wood, concrete 
or ceramic tile. If 
floors are 
covered with 
synthetic 
material, the 
relative humidity 
should be at least 
30 %. 

Power frequency 
(50/60 Hz) 
magnetic field 
IEC 61000-4-8 

3 A/m 

3 A/m 

Power frequency 
magnetic fields 
should be at 
levels 
characteristic of a 
typical location in 
a typical 
commercial or 
hospital 
environment. 

 

Table 3 

For ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING 

Guidance and manufacturer’s declaration - electromagnetic immunity 

The TMA-6000 is intended for use in the electromagnetic environment specified below. The 
customer or the user of the TMA-6000 should assure that it is used in such an environment.

IMMUNITY test 

IEC 60601test level 

Compliance 

level 

Electromagnetic environment 

- guidance 

 
 
 
 
 
 
 
 

 
 
 
 
 
 
 
 

 
 
 
 
 
 
 
 

Portable and mobile RF 
communications equipment 
should be used no closer to any 
part of the TMA-6000, including 
cables, than the recommended 
separation distance calculated 
from the equation applicable to 
the frequency of the transmitter.

Summary of Contents for TMA-6000

Page 1: ...TIONS 3 NOTICE 4 SETUP AND OPERATING PROCEDURES 5 1 BATTERY LOADING 5 2 CLOCK AND DATE ADJUSTMENT 6 3 CONNECTING THE CUFF TO THE MONITOR 6 4 APPLYING THE CUFF 7 5 BODY POSTURE DURING MEASUREMENT 7 6 T...

Page 2: ...usually appears in the daytime and lowest one usually at midnight Typically the value begins to increase at around 3 00AM and reaches to highest level in the daytime while most people are awake and ac...

Page 3: ...lectrical equipment Part 1 General requirements for basic safety and essential performance IEC60601 1 2 2007 EN 60601 1 2 2007 AC 2010 Medical electrical equipment Part 1 2 General requirements for ba...

Page 4: ...o allow the blood circulation in your arm to recover Prolonged over inflation cuff pressure exceed 300 mmHg or maintained above15 mmHg for longer than 3 minutes of the bladder may cause ecchymoma of y...

Page 5: ...nd complied with IEC 60601 1 EN 60601 1 UL 60601 1 and IEC 60601 1 2 EN 60601 1 2 UL 60601 1 2 is suitable for this monitor Please note that the monitor jack size hole 3 5mm center pin 1 0mm Please pa...

Page 6: ...ode press the START and MEM button simultaneously for two seconds the month will blink at first Press the button START repeatedly the day hour and minute will blink in turn While the number is blinkin...

Page 7: ...PECIFICATIONS to make sure that the appropriate cuff is used 2 Measuring on same arm each time 3 Do not move your arm body or the monitor and do not move the rubber tube during measurement 4 Stay quie...

Page 8: ...rtbeat symbol if any will blink on the screen The result will be automatically stored in the monitor d After measurement the monitor will turn off automatically after 1 minute of no operation Alternat...

Page 9: ...ntended to provide a basis of any type of rush toward emergency conditions diagnosis based on the color scheme and that the color scheme is meant only to discriminate between the different levels of b...

Page 10: ...ened Apply the cuff correctly and try again Body posture was not correct during testing Review the body posture and testing sections of the instructions and re test LCD Display shows abnormal result S...

Page 11: ...chnical personnel to repair those parts of equipment which are designated repairably can be supplied 9 The monitor can maintain the safety and performance characteristics for a minimum of 10 000 measu...

Page 12: ...magnetic emissions The TMA 6000 is intended for use in the electromagnetic environment specified below The customer or the user of the TMA 6000 should assure that it is used in such an environment Emi...

Page 13: ...requency 50 60 Hz magnetic field IEC 61000 4 8 3 A m 3 A m Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment Tabl...

Page 14: ...tions Electromagnetic propagation is affected by absorption and reflection from structures objects and people a Field strengths from fixed transmitters such as base stations for radio cellular cordles...

Page 15: ...istance according to frequency of transmitter m Rated maximum output power of transmitter W 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 GHz 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3...

Reviews: