Calibration
Remove the T-Micro CO2 controller from the wall
socket.
Remove the plug from the CO2-source from the
T-Micro CO2 controller.
Make sure that a CO2-sensor is connected to one
of the two connection outlets.
Plug the T-Micro CO2 controller back in the wall
socket.
During start up, push the button top left with a pen
or paperclip once.
After start up (takes approx. 12 seconds) the
calibration routine will start (the display shows
“cal! 8” and will countdown from 8 to 0).
Cal! 6
The display shows “Start?”
Start?
Again, push the button top left and hold until
“Starting” appears on the display.
Starting
(takes approx. 8 seconds, in the meantime a
growing arrow will appear on the screen).
->
The calibration routine has now started and will
take approx. 20 minutes.
*******
After calibration, put DIP-switch 1 back to
position 0.
Close the sensor case.
TECHNICAL SPECIFICATIONS:
Power supply:
100 to 240 Volt
Net frequency:
50 to 60 Hz
Consumption:
3 Watt (with sensor, without CO2-source)
Maximum load:
1000W (max 4A)
Adjustment range CO
2
set:
300 to 2000 ppm
Adjustment range CO
2
+/-:
20 to 200 ppm (advice 50 ppm)
(hysteresis/bandwith)
Measurement frequency:
1 measurement per second