CHAPTER 1 INSTALLATION
14
1.3 Installation Environment
Work environment will have great influence on instrument performance and life
time. Before installing UV2600, confirm the following environment conditions
at installation site.
1.3.1 Operating Temperature
Allowed environmental temperature range
:
5
~
35
℃
.
However, it is suggested
to install air conditioner and set room temperature to 20
~
25
℃
for instrument
under stable work status.
1.3.2 Operating Humidity
Allowed environmental humidity range
:
45
%~
85
%。
It is suggested to install
dehumidifying device and control the hunidity in room within 30
%~
70
%
for
instrument under stable work status.
1.3.3 Storage Temperature
Allowed temperature range for storage
:-
20
~
60
℃
。
1.3.4 Atmospheric Environment
1
)
Free from acid gas or alkaline gas and other gases which may corrode
metals significantly.
2
)
Free from gases which may dissolve paint such as from organic solvents
(particularly benzene, thinner etc).
1.3.5 Sample Solvent
1
)
Sample solvent and corrosive material should not to be put near instrument.
When using them, be careful to avoid drop or spill onto the cover or inside
sample compartment.
2
)
If organic solvent has been spilled on the cover or inside sample
compartment, please cleans it by soft cloth or paper at once.
3
)
If acid or alkali solution has been spilled inside sample compartment, please
take apart the compartment to wash and clean.
1.3.6 Other General Notes
a) Avoid direct sunlight (otherwise optical performance might deteriorate or the
housing might become discolored). Avoid installation by a window if at all
possible.
b) Vibrations or shocks strong enough to be felt by the human body must not
be transmitted to the instrument (otherwise the fine adjustment mechanism
might malfunction).
c) Avoid installation in near heat generating apparatus such as a gas burner,
electric heater or oven in order to prevent the mainframe cover being heated
beyond 70 Deg C.
d) Avoid installation near instruments which generate a strong electric field