2 - Hardware Setup
Settings and Options
Cavro® XLP 6000 Modular Syringe Pump Operating Manual, 734237-C, en
2 - 9
2.4
Settings and Options
2.4.1
Configuration Commands
The XLP 6000 firmware allows the user to configure the pump for different modes
of operation. The U commands (see Chapter 3, "Software Communication") are
used to write the configuration information to the non-volatile memory and control
the following options:
Valve type:
The pump can be configured to operate with different valve
options (3-port, 4-port, T-valve, Y Block, 3-port distribution, 6-port distribution,
and 9-port distribution).
Baud rate:
RS-232/RS-485 communication is possible at 9600 baud (default)
and 38400 baud. CAN communication is possible at 100K baud (default),
125K baud, 250K baud, 500K baud, and 1M baud rates.
Non-Volatile Memory Auto Mode:
Allows the pump to run command strings
out of the non-volatile memory.
2.4.2
Address Switch Settings
The address switch (see Figure 2-5, "Address Switch") is located near the top of
the XLP 6000 electronic circuit board. It is used to give each XLP 6000 in a multi-
pump configuration a unique or specific address, allowing the user to direct
commands to specific pumps. The address switch has sixteen positions
(numbered 0 through F). Fifteen positions (addresses 0 through E) are valid pump
addresses.