2 - 10
Cavro® XLP 6000 Modular Syringe Pump Operating Manual, 734237-C, en
2 - Hardware Setup
Settings and Options
Figure 2-5
Address Switch
To set the address switch:
To set the address switch, use a jeweler’s screwdriver or small flat head
screwdriver and turn the switch in either direction to the desired position.
Note: Power cycle (or power up) the pump after setting the address switch.
For information on the addressing schemes for different pump configurations, see
Chapter 3, "Software Communication".
2.4.3
Self-Test
The “F” address switch position is used to activate the XLP 6000 self-test. Self-
test causes the XLP 6000 to initialize, then cycle repeatedly through a series of
plunger movements. The self-test cycles through speed codes 0 to 14. If an error
condition occurs, the pump stops moving.
To run the self-test, set the address switch to position “F.” Then supply power to
the pump.
Caution! Always run liquid through the syringe and valve. Failure to do so can
damage the valve and syringe seal.