WARRANTY & REPAIRS 

All units carry a warranty against manufacturing defects for a period of 3 years from the date of 

purchase. If for any reason you require your unit to be serviced or repaired the repair 

must 

be 

carried out by a Brymill Authorised Repair Center. 

To claim warranty the unit must be sent back to where it was originally purchased from. 

The warranty does not cover damage as a result of misuse and abuse. 

If repairs are performed by any other party the warranty will become invalid. Unauthorized 

repairing will also absolve Brymill Cryogenic Systems of any claims for injury caused by an 

unauthorized repaired unit. 

Warranty on Replacement Flask is 3 months. 

NOTES & TIPS

Caution – When handling Liquid Nitrogen ensure you are familiar with the information 
contained in the Material Safety Data Sheet for Liquid Nitrogen and that you are wearing the 
appropriate recommended Personal Protective Equipment. 

The Cryosurgical Unit is easily  ed warm or re led cold a er prolonged use. 

Liquid Nitrogen may be carefully poured into the Bo le (slowly when warm) or by any standard 

LOW pressure withdrawal device from a Liquid Nitrogen storage Dewar. 

The volume of Liquid Nitrogen needed for adequate func oning is from 33% to 70% full. 

It is recommended that for a 3 - 6 hour dura on of intermi ent use that the Cryosurgical Unit 

be 70%  ed. 

However, the physician will  nd that if the unit is  ed approximately 40%, it will be lighter in 

weight and the Top will remain slightly warmer for further comfort in con nued handling. 

Before replacing the top, ensure that the rubber gasket is s ll in place inside the cap. If it is 

missing the Cryosurgical Unit may not pressurize correctly and the top may become stuck. In 

this event the Cryosurgical Unit must be returned to an Authorized Repair Centre for proper 

removal. 

A er  lling a warm Cryosurgical Unit, allow 30 to 60 seconds for the ini l boiling of the Liquid 

Nitrogen to subside before a emp ng to replace the Top. If a large number of cryosurgery 

procedures are scheduled, the reservoir of Liquid Nitrogen may be topped o  a er the  t 

boiling, and the unit has cooled o . 

Caution - To re l a Cryosurgical Unit after it has been in use you must ensure that the unit is
depressurized before removing the top. To Depressurize the Cryosurgical Unit, unscrew the 
top a quarter to half turn only. The pressurized gas inside will begin to vent from the hole 
situated in front of the Valve Body. Once the hissing has stopped the top can be unscrewed 
and removed. 