TROUBLESHOOTING GUIDE 

ISSUE 

SOLUTION 

If the Cryosurgical Unit 
does not spray or sprays 
only intermittently. 

Spray Tip may be clogged.  Remove  p.  If the Cryosurgical Unit sprays

without a  p, clean the opening of the  p with a  ne needle or bang the  p 

gently on a table or counter to dislodge any foreign ma er. Then check the 

Liquid Nitrogen supply (Dewar) for contaminants that can clog the  ps and 

unit. To help ensure the Liquid Nitrogen remains free of par

late ma er, 

such as ice crystals, carbon dioxide slush, lint, etc., the storage Dewar used 

should be completely emp d at least 4  mes a year just prior to having it 

lled. This is accomplished by vigorously agita ng the residual amount of 

Liquid Nitrogen in the Dewar and discarding it in a safe, outdoor area. 

If the unit s l has a problem it may have been over  lled and there is 

insu ient air space inside the bo le to create an adequate build up of 

pressure to enable that is required for the Liquid Nitrogen to spray. 

Check that the Gasket is in place inside the cover and is not split or missing. 

Always ensure you have spare Gaskets available. 

Trigger Handle sticks 
open. 

Valve Stem s cking. Depressurize the unit immediately by unscrewing the 

top a quarter to one half turn. Lubricate valve stem as follows: 

Wait un  the Cryosurgical Unit is warm and dry, the top center valve stem 

should be lubricated with a 

DROP 

of silicone lubricant or WD-40. 

Lub

on should be carried out every 3 to 6 months. 

CAUTION: 

If an excessive amount of lubricant is applied the trigger 

mechanism could freeze open. 

Unit appears to be 
“Leaking” or “Hissing” 
from the Relief Valve.  

During normal op

g cond ons, if the Unit is le  standing for a period 

of  me the constant evapo

on of the Liquid Nitrogen inside the bo le 

will result in the temporary opening of the Relief Value v

ng the excess 

pressure. This “hissing” is also heard when the Unit is picked up. 

THIS IS 

NOT A PROBLEM

. The Relief Valve is just op

g as designed. 

Unit is frosting over and 
excessive hissing from 
the Relief Valve. 

If the Exterior of Bo

 is fro

g over there will be excessive Relief Valve 

vity (hissing). 

THIS IS A PROBLEM

. The vacuum inside the bo le has 

deteriorated due to age or the bo le has been damaged. 

Replacement of Bo le is necessary. Return the complete Cryosurgical Unit 

to an Authorized Brymill Dealer.   

Note:   If the unit is found to have been misused or abused  

(Dropped, dented, cracked etc.) the 3 year warranty will be void. 

Trigger does not operate. 

The posi on of the Trigger should be on the le  or right hand side of the 

delivery tube and not at the back of the Unit. 

If you experience any further problems with your Cryosurgical Unit contact the 

company where you purchased it immediately. 