TROUBLESHOOTING GUIDE
ISSUE
SOLUTION
If the Cryosurgical Unit
does not spray or sprays
only intermittently.
Spray Tip may be clogged. Remove p. If the Cryosurgical Unit sprays
without a p, clean the opening of the p with a ne needle or bang the p
gently on a table or counter to dislodge any foreign ma er. Then check the
Liquid Nitrogen supply (Dewar) for contaminants that can clog the ps and
unit. To help ensure the Liquid Nitrogen remains free of par
late ma er,
such as ice crystals, carbon dioxide slush, lint, etc., the storage Dewar used
should be completely emp d at least 4 mes a year just prior to having it
lled. This is accomplished by vigorously agita ng the residual amount of
Liquid Nitrogen in the Dewar and discarding it in a safe, outdoor area.
If the unit s l has a problem it may have been over lled and there is
insu ient air space inside the bo le to create an adequate build up of
pressure to enable that is required for the Liquid Nitrogen to spray.
Check that the Gasket is in place inside the cover and is not split or missing.
Always ensure you have spare Gaskets available.
Trigger Handle sticks
open.
Valve Stem s cking. Depressurize the unit immediately by unscrewing the
top a quarter to one half turn. Lubricate valve stem as follows:
Wait un the Cryosurgical Unit is warm and dry, the top center valve stem
should be lubricated with a
DROP
of silicone lubricant or WD-40.
Lub
on should be carried out every 3 to 6 months.
CAUTION:
If an excessive amount of lubricant is applied the trigger
mechanism could freeze open.
Unit appears to be
“Leaking” or “Hissing”
from the Relief Valve.
During normal op
g cond ons, if the Unit is le standing for a period
of me the constant evapo
on of the Liquid Nitrogen inside the bo le
will result in the temporary opening of the Relief Value v
ng the excess
pressure. This “hissing” is also heard when the Unit is picked up.
THIS IS
NOT A PROBLEM
. The Relief Valve is just op
g as designed.
Unit is frosting over and
excessive hissing from
the Relief Valve.
If the Exterior of Bo
is fro
g over there will be excessive Relief Valve
vity (hissing).
THIS IS A PROBLEM
. The vacuum inside the bo le has
deteriorated due to age or the bo le has been damaged.
Replacement of Bo le is necessary. Return the complete Cryosurgical Unit
to an Authorized Brymill Dealer.
Note: If the unit is found to have been misused or abused
(Dropped, dented, cracked etc.) the 3 year warranty will be void.
Trigger does not operate.
The posi on of the Trigger should be on the le or right hand side of the
delivery tube and not at the back of the Unit.
If you experience any further problems with your Cryosurgical Unit contact the
company where you purchased it immediately.