background image

Precautions when placing loads on the 

unit

 

o

Make sure the load is perpendicular to the surface to 
which this unit is attached.

 

o

Apply the load so that it is centered on the center of 
the unit. If the load is not centered (eccentric load), 
twisting, for example, and measurement errors could 
occur. This could even result in damage.

F1

F2

F

F is the correct load orientation
F1 is an eccentric inclined load
F2 is an eccentric load

 

o

If the load receiving area is contacted by something that 
is at a different temperature and the load is increased, 
the values output by this device could become unstable, 
making accurate measurement impossible. In such a 
case, wait until the temperature difference ceases to 
exist before measuring.

Electrical connection of load cell

 

o

Connect as shown in the illustration below. Incorrect 
connections could result in inability to balance and 
in errors occurring in the output voltage when loads 
are applied.

Using a cable with bare lead wires

Red 

Input (+)

Black 

Output (−)

Blue 

Input (−)

White 

Output (+)

Yellow Shield

 

o

The shield is not connected to the main body of this 
product. For this reason, if grounding is necessary because 
of external noise, for example, arrange to ground the 
shield to a part other than the body of this unit.

 

o

Since the cable is directly connected to this unit, use 
a specialized cable to increase the length. (Please 
consult with us.)

Using a connector (optional)

A  Input (+)

B  Output (−)

C  Input (−)

D  Output (+)

E Shield

NDIS-compatible  

connector

Handling after use

 

o

When moving this unit while it is installed, take protective 
measures to prevent it from being shaken or subjected 
to excessive external forces.

 

o

When storing it, keep it in a dry place where it will not 
be exposed to water or oil, for example.

Specifications

TU-PGRS 100N-G to TU-PGRS 20KN-G
Rated capacity:  100 N, 200 N, 500 N, 1 kN, 2 kN, 3 kN, 5 kN, 

10 kN, 20 kN

Safe overload rating: 150% R.C.
Rated output: 2 mV/V ±0.3%
Linearity: 0.03% R.O.
Hysteresis: 0.03% R.O.
Repeatability: 0.02% R.O.
Zero balance: 3% R.O.
Safe excitation voltage: 15 V (100 N, 200 N)
 

   

20 V (500 N – 20 kN)

Input terminal resistance: 1050 ±10 Ω
Output terminal resistance: 1050 ±10 Ω
Insulation resistance: 1000 MΩ or higher (DC 50 V)
Compensated temperature range: −10 to 60°C
Permissible temperature range: −30 to 80°C
Temperature effect on zero balance: 0.025% R.O./10°C
Temperature effect on output: 0.03% R.C./10°C
Cable:  Ø6mm 4-core shielded, 5m direct connection with 

bare lead wires (100 N – 5 kN)

 

  Ø8mm 4-core shielded, 5m direct connection with 

bare lead wires (10 kN, 20 kN)

Body material: aluminum (coated surface) (100 N–5 kN)
 

Steel (coated surface) (10 kN, 20 kN)

Environmental compliance: RoHS (10 substances)
TEDS: not supported
Other: load button included

Dimensional drawings 

(dimensions in millimeters (mm))

A

(Ø14)

Ø58

Ø11

Ø44

Ø11

Ø44

26

B

Rubber sleeve

Cable

M8X1.25

Ø51

3x Ø4.5 thr

ough holes

Rated capacity

A

B

100 N

0.15

13

200 N

0.15

13

500 N

0.09 11.5

1 kN

0.09 11.5

TU-PGRS 2K-G to TU-PGRS 5KN-G

(Ø18)

Ø88

Ø62

Ø21

36

18

Rubber sleeve

Cable

M12X1.75

Ø75

6x Ø6.5 thr

ough holes

Ø21

Ø62

0.08

TU-PGRS 10K-G to TU-PGRS 20KN-G

Ø118

Ø79

Ø34

40

20

Rubber sleeve

Cable

M18X1.5

Ø98

6x Ø9 thr

ough holes

0.08

Ø34

Ø79

R1/2 (PT1/2)

Warranty explanation

 

o

The warranty period for this device is one year from 
the date of purchase.
1)  Be aware that repairs will require payment in the 

following cases even during the warranty period. 
Malfunction or damage due to misuse

2)  Malfunction or damage caused by modifications or 

repairs conducted by any party other than our com-
pany or a service person designated by our company

3)  Malfunction or damage caused by dropping, trans-

portation or similar handling after product delivery

4)  Malfunction or damage caused by fire, earthquake, 

water, lightning or other natural disaster

5)  Malfunction or damage caused by external fac-

tors, including power supplies and equipment 

environmental conditions, that deviate from the 
operation requirements of this product

6)  Malfunction or damage if the product was not pur-

chased from our company or an agent designated 
by our company

 

o

We offer paid service after the conclusion of the warranty 
period. For details, please contact the retailer where you 
purchased the unit.

 

o

Be aware that our company will bear no responsibility 
for any secondary damages resulting from the operation 
of this device or related to data.

 

o

Information is given about products in this manual 
only for the purpose of example and does not indicate 
any guarantees against infringements of third-party 
intellectual property rights and other rights related to 
them. TEAC Corporation will bear no responsibility for 
infringements on third-party intellectual property rights 
or their occurrence because of the use of these products.

Contact information

TEAC CORPORATION (Manufacturer)

1-47 Ochiai, Tama-shi, Tokyo 206-8530 Japan
Phone: +81-042-356-9154 

TEAC AMERICA, INC.

10410 Pioneer Blvd. Suite #1, Santa Fe Springs, California 90670, 
U.S.A.
Phone: +1-323-726-0303

TEAC EUROPE GmbH. (EU Importer)

Bahnstrasse 12, 65205 Wiesbaden-Erbenheim, Germany
Phone: +49-611-7158-349

0720 MA-2733A

TU-PGRS 100N-G to TU-PGRS 1KN-G