strain by referencing the material in the back of this manual, then make a
zero adjustment.
■ Peak Hold Display
Press ⑫ PEAK HOLD button and the display shows [P] then the peak hold
operation starts. It continues until ⑫ PEAK HOLD button is pressed again.
7.3.2 User Unit Display Measurement
Connect ⑥ INPUT in the rear panel to a sensor, switch ④ INDICATION to USER,
and the display mode becomes a user display mode. The displayed value is computed
according to the user defined physical quantity (Kg, etc.) vs. sensor
sensitivity (mV/V) setting.
■ Sensor Sensitivity (mV/V and rated capacity) setting procedure
Press ⑪ ENTER while holding ⑬ SELECT in the operation panel, and the
display changes to C 00.000. Move to a desired digit using ⑧(
→)
or ⑩(
←)
key, and select values using ⑦(↑) or ⑨(↓) key. Once a desired value is
set, press ⑪ ENTER. Then the rated capacity is displayed as G 00000. Move
to a desired digit using ⑧
(→)
or ⑩
(←)
key, and select values using ⑦
(↑) or ⑨
(↓)
key. (The rated capacity is a value found in the sensor
s
data sheet)
To change the decimal point position, move the cursor to a desired decimal
point position using ⑧
(→)
or ⑩
(←)
key, press ⑪ ENTER key, and the
character display becomes reversed. Move to a desired position using ⑧
(→)
or ⑩
(←)
key, and press ⑪ ENTER. When all of the settings are finished,
press ⑪ ENTER one more time.
■ How to set the number of digits below decimal point in the measurement screen
The number of display digits below decimal point can be selected during
measurement. Press ⑧(
→
)or ⑩(
←
) key while holding ⑬ SELECT, and the
character at the cursor becomes inverted. Move to a desired digit using ⑧
(
→
) or ⑩(
←
) key, then press ⑪ENTER. Note that the maximum value to be
displayed is 9999 including the number after the decimal point.
■ Zero adjustment and peak hold display
The operation and function are the same as strain measurement. Switch ⑤ ZERO
SET in the rear panel to "ON". The zero adjustment is engaged by pressing
⑨
(↓
key
)
ZERO RESET while holding ⑬ SELECT button. The adjustable range
is ±10000×10
−6
in strain quantity.
■ Display accuracy
The numerical value that is smaller than 1×10
−6
strain in
input