4.CALIBRATION
24
・
Equivalent Input Calibration
If a calibration error display appears, take a proper action according to the error,
then perform calibration again.
…
Span set value is "00000". Set a correct span value.
On completion of calibration, turn on the calibration LOCK.
1) Start the equivalent input calibration.
2) Set the rated output value of the sensor.
Use
and
keys
to set the rated output value,
then use
key to validata the setting.
3) Set the rated value.
Use
and
keys to set the rated value,
then use
key to validate the setting. Press
key at a blinking
digit to place a decimal point immediately after the digit.
4) Place the sensor without load and set the zero point. Check that the sensor is
unloaded,then press
and
keys in this order.
If the indicated value following the display CAL2E is
zero, the equivalent input calibration is terminated.
SHIFT
CAL
ENT
UP
DOWN
SEL
ENT
UP
DOWN
SEL
ENT
FNC
ZERO
ENT
Equivalent Input Calibration
Rated Output Value
(
0.500 to 3.000mV/V
)
Rated Value
(
00000 to 19999
)
Example of a sensor having the output
value of 2.010mV/V for rated 100.00kgf
(
N
)
c