Teac TC-MFSR(T)-G Instructions For Use Download Page 2

of external noise, for example, arrange to ground the 
shield to a part other than the body of this unit.

 

o

Since the cable is directly connected to this unit, use 
a specialized cable to increase the length. (Please 
consult with us.)

 

o

When conducting insulation resistance tests, limit them 
to the red, black, blue and white cores. Do not apply to 
the TEDS cores (orange and green).

Using a connector (optional)

Input (+)

Output (−)

Input (−)

Output (+)

TEDS signal (+)

TEDS signal (COM)

E Shield

NDIS-compatible connector

TEDS

TEDS overview

TEDS (Transducer Electronic Data Sheet) is a memory chip 
that can electronically read and write sensor-specific data. 
The TEDS built into this unit has serial number, load cell rated 
output and manufacturer name data recorded on it. By 
connecting an indicator that supports TEDS, the TEDS data 
of the connected load cell will be automatically read and 
equivalent input calibration will be completed. (For details, 

read the operation manual of the connected indicator.)
Refer to the connection diagrams in the “Electrical con-
nection of load cell with built-in TEDS”
 section as well as 
the operation manual for the indicator being connected 
for procedures to make connections.
TEAC indicators that support TEDS include the TD-01 Portable, 
TD-700T, TD-260T, TD-275T and TD-280T. For details, inquire 
at a shop that handles these products.
If not using the TEDS function, do not connect the orange 
and green load cell cores.
Moreover, take measures to prevent the orange and green 
cores from touching other terminals.
Data recorded in TEDS are tested values from calibration 
conducted at room temperature during inspection before 
shipping from our company.
Load cell output will be affected by the environmental 
temperature where used, even when within the compen-
sated temperature range established in the specifications. 
Although load cell output is calculated from the calibration value 
saved in TEDS, when the environmental temperature differs 
greatly from room temperature, the temperature impacts increase 
on the zero point and the output. For this reason, consideration 
of the effect on output voltage is necessary. Output voltage 
changes will be within the specification range as long as the 
unit is used at temperatures within the specification range.

When using the TEDS function

This product’s cores for TEDS (orange and green) are 
protected by a plastic cover as shown in the illustration.
This is to prevent miswiring and contact with other connec-
tors during use, for example, which could cause shorting.
To use the TEDS function, remove the tip of the plastic cover, 
following the perforations, from the end of the cables. Then, 
connect the TEDS cores (orange and green) to the indicator.
Refer to the “Electrical connection of load cell with built-in 
TEDS”
 section for how to connect each wire.

Cores for sensor

Load cell cable

Protective 

plastic

Perforations

Cores for TEDS

(orange and green)

When not using the TEDS function

If not using the TEDS function, remove the entire protective 
plastic cover and cut off the TEDS cores (orange and green) 
around where the plastic cover was attached to the cable.
If you use it without removing these cores, do not remove 
the protective plastic or make other arrangements to prevent 
the cores from touching other places.
Refer to the “Electrical connection of load cell with built-in 
TEDS”
 section for how to connect each wire.

Note about the tag attached to the TEDS 
cable

Since the protective plastic and tag are not suitable for 
the entire temperature range of this unit’s specifications, 
do not expose them to high temperatures.

Handling after use

 

o

When moving this unit while it is installed, take protective 
measures to prevent it from being shaken or subjected 
to excessive external forces.

 

o

When storing it, keep it in a dry place where it will not 
be exposed to water or oil, for example.

Specifications

Rated capacity: 20 N, 50 N
Safe overload rating: 120% R.C.
Rated output: about 1 mV/V
Linearity: 0.5% R.O.
Hysteresis: 0.5% R.O.
Repeatability: 0.3% R.O.
Zero balance: ±10% R.O.
Safe excitation voltage: 6 V
Input terminal resistance: 350 ±20 Ω
Output terminal resistance: 350 ±20 Ω
Insulation resistance: 1000 MΩ
Compensated temperature range: 0 to 60°C
Permissible temperature range: −10 to 80°C
Temperature effect on zero balance: 2% R.O./10°C
Temperature effect on output: 1% R.C./10°C
Cable: Ø2mm 6-core robot cable
 

3 m with bare lead wires

Body material: beryllium copper
Environmental compliance: RoHS (10 substances)
Other: built-in TEDS

 

o

Since the cable tip is not compatible with the entire 
temperature range of this unit’s specifications, use it 
in a room-temperature environment.

Dimensional drawing

TC-MFSR(T)-20N-G, TC-MFSR(T)-50N-G

Dimensions in millimeters (mm)

Ø8

Ø12

(5)

4.55

3.35

0.5

(Ø2.3)

Ø2 cable

1.55

Warranty explanation

 

o

The warranty period for this device is one year from 
the date of purchase.
Be aware that repairs will require payment in the following 
cases even during the warranty period.
1)  Malfunction or damage due to misuse
2)  Malfunction or damage caused by modifications or 

repairs conducted by any party other than our com-
pany or a service person designated by our company

3)  Malfunction or damage caused by dropping, trans-

portation or similar handling after product delivery

4)  Malfunction or damage caused by fire, earthquake, 

water, lightning or other natural disaster

5)  Malfunction or damage caused by external factors, 

including power supplies and equipment environ-
mental conditions, that deviate from the operation 
requirements of this product

6)  Malfunction or damage if the product was not pur-

chased from our company or an agent designated 
by our company

 

o

We offer paid service after the conclusion of the warranty 
period. For details, please contact the retailer where you 
purchased the unit.

 

o

Be aware that our company will bear no responsibility 
for any secondary damages resulting from the operation 
of this device or related to data.

 

o

Information is given about products in this manual 
only for the purpose of example and does not indicate 
any guarantees against infringements of third-party 
intellectual property rights and other rights related to 
them. TEAC Corporation will bear no responsibility for 
infringements on third-party intellectual property rights 
or their occurrence because of the use of these products.

Contact information

TEAC CORPORATION (Manufacturer)

1-47 Ochiai, Tama-shi, Tokyo 206-8530 Japan
Phone: +81-042-356-9154 

TEAC AMERICA, INC.

10410 Pioneer Blvd. Suite #1 and #4, Santa Fe Springs, 
California 90670, U.S.A.
Phone: +1-323-726-0303

TEAC EUROPE GmbH. (EU Importer)

Bahnstrasse 12, 65205 Wiesbaden-Erbenheim, Germany
Phone: +49-611-7158-349

0719 MA-2601A

Reviews: