5-6. Storing Data to PC
- CA-20 can store measured sensitivity data to pc using data transfer application supplied with CA-20.
- Please refer to application manual for the details.
6. Specification
Charge type accelerometers 0.255 to 196 pC/G
Calibration range
IEPE type accelerometers 0.255 to 196 mV/G
Calibratable sensor weight
Approx. 100g or less
Calibration accuracy
±(3%+2digit)(23±5ºC)
Exciter vibration acceleration
1G(rms) ±3%(23±5ºC)
Exciter vibration frequency
159.2Hz±2%
Sensor attachment
M6 female, depth of 6.5 mm
Operating temperature range
10ºC to 40ºC
Storage temperature range
-10ºC to 50ºC
Operating Altitude
Max.2000 meters
Batteries
LR-6 x 4 batteries (5V DC, 0.75A )
Continuous use: approx. 1 hours
(Calibration sensor weight: approx 25g )
Battery life
(LCD ON)
Power supply for accelerometer
Selectable DC 24 / 15 V, 0.5 / 2 / 4mA Constant current supply
AC Adapter (AC100-240V 50/60Hz 0.3A)
Power supply for unit
To be used attached AC Adapter with CE Marking
Dimensions(W×H×D)
Approx. 120 mm × 50 mm × 140 mm
Weight
Approx. 1 kg (Including batteries)
Accessories Included Cable
LN-030 0.5m Miniature-BNC 1
Conversion studs
M6-M3, M6-10-32, M6-flat, M6 standard stud (one each)
Packing box
1
Instruction manual
1
11
Summary of Contents for CA-20