Special Instructions for medical applications (IEC/EN/UL/CSA 60601-1)

These products are designed for continuous operation within an overall enclosure, and must be mounted

such that access to the mains terminals is restricted. See clause 16, IEC60601-1, EN60601-1 and UL/CSA

60601-1.
These products are NOT suitable for use in the presence of flammable anaesthetic mixtures with air or with

oxygen, or with nitrous oxide.
These products are classed as ordinary equipment, type B and have a Reinforced barrier between input and

output to IEC60601-1, EN60601-1 and UL/CSA 60601-1. They are NOT protected against the ingress of

water.
All outputs have basic spacings to earth rated for mains - 250Vac, and due consideration must be given to

this in the end product design.
Reference should be made to local regulations concerning the disposal of these products at the end of their

useful life.

Environmental Specifications:

Description

Operation

Storage

Use

Indoor

-

Temperature

0°C  -  +70°C (derating at 2.5%/°C above 50°C)

-40°C  -  +85°C

Humidity

5  -  95% RH, non-condensing

5  -  95% RH, non-condensing

Altitude

-200m  -  3000m

-200m  -  5000m

Pressure

70kPa  -  106kPa

54kPa  -  106kPa

Orientation

ALL

Shock

+ / - 3 x 30G shocks in each plane, total 18 shocks. 30G shocks are 11ms (±0.5ms),

half sine. Conforms to EN60068-2-27, EN60068-2-47, IEC68-2-27, IEC68-2-47, JIS

C0041-1987.

Vibration

Single axis 10 - 500Hz at 2G (sweep and endurance at resonance) in all 3 planes.

Material Group

IIIb

Pollution Degree

2

Overvoltage Category

II

Class

I

Level of Insulation:

Dielectric Strength testing is carried out as follows:

Primary mains circuit to earth: - 2.25 - 2.35kVDC

Primary mains circuits to transformer core: - 4.25 - 4.35kVDC*

Primary mains circuits to secondary: -4.25 - 4.35kVDC*

Outputs to each other and to earth are isolated to XXXVDC.

*This test is not possible with Y capacitors fitted to the unit as damage to these

capacitors will occur. It is also necessary to short circuit the outputs together and to earth.

Summary of Contents for CSS65

Page 1: ...it Output Loading The output power taken from the power supply must not exceed the rating stated on the power supply label except as stated in the product limitations in this handbook Input Parameters...

Page 2: ...have an obligation to protect service engineers as well as users In order to comply with this a label must be fitted to these products which is clearly visible to service personnel accessing the overa...

Page 3: ...Ausgangsstrom des Netzteiles darf die Leistung die auf dem Label des Netzteiles vermerkt ist nur dann berschreiten wenn dies in den Produktgrenzen dieses Handbuches ausgezeichnet ist Eingangsparamete...

Page 4: ...ller f r den Schutz von Wartungspersonal und Benutzern verantwortlich Um diesen Bestimmungen gerecht zu werden muss auf den Produkten ein Label angebracht werden das deutlich sichtbar f r das Wartungs...

Page 5: ...sit soutir e L intensit soutir e de l alimentation ne doit pas d passer l intensit nominale marqu e sur la plaque signal tique sauf indications contraires dans les limitations du produit dans ce manue...

Page 6: ...ger les techniciens d entretien et les utilisateurs Pour cela une plaque signal tique doit tre install e sur ces produits et cette plaque doit tre bien visible pour les techniciens d entretien interve...

Page 7: ...ruppo IIIb Carico in uscita La potenza in uscita ottenuta dall alimentatore non deve superare la potenza nominale indicata sulla targhetta dell alimentatore fatto salvo dove indicato nei limiti per i...

Page 8: ...i tecnici di manutenzione e inoltre gli utenti Per far fronte a tali obblighi i prodotti devono presentare una targhetta chiaramente visibile al personale di manutenzione che accede all apparecchiatu...

Page 9: ...ia de salida tomada de la fuente de alimentaci n no puede sobrepasar el valor nominal indicado en la etiqueta de la fuente de alimentaci n excepto en los casos indicados en las limitaciones del produc...

Page 10: ...teger a los ingenieros de servicio adem s de a los usuarios Para que esto se cumpla debe colocarse una etiqueta en estos productos que pueda ser vista claramente por el personal de servicio que accede...

Page 11: ...da A pot ncia de sa da extra da da fonte de alimenta o n o deve exceder a classifica o assinalada na etiqueta da fonte de alimenta o excepto quando indicado nas limita es do produto neste guia Par met...

Page 12: ...teger os t cnicos de manuten o bem como os utilizadores De forma a respeitar este regulamento estes produtos dever o ter uma etiqueta que seja facilmente vis vel ao pessoal de assist ncia que aceda ao...

Page 13: ...end of their useful life Environmental Specifications Description Operation Storage Use Indoor Temperature 0 C 70 C derating at 2 5 C above 50 C 40 C 85 C Humidity 5 95 RH non condensing 5 95 RH non c...

Page 14: ...h Molex Housing 09 50 3041 All individual outputs are SELV Note If the total voltage of outputs connected in series exceed the 60Vdc SELV limit then all outputs must be considered non SELV Non SELV ou...

Page 15: ...id as flat to the earthed metal box plane as possible All output cables should be routed as far away from input cables as possible If the input and output cables must be run close to each other then s...

Page 16: ...tive earth in the final system assembly Fusing Internal fuses F1 F2 The product is dual fused in the L and N line T3 15A 250V The product is not user serviceable Fuses must not be replaced by the user...

Reviews: