1. Product Specifications

Product Model . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . WB-300 (Basic Electronic Medical Beam Scale)

WB-300MH (Electronic Medical Beam Scale

with Mechanical Height Rod)

WB-300DH (Electronic Medical Beam Scale 

with Digital Height Rod and BMI Function)

Type of Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Four (4) Half Bridge Load Cell
Capacity  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .440 lb – 0.1 lb

200 kg – 0.05 kg

Graduation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0.1 lb, 1/4 lb, 0.05 kg
Range of Height (WB-300MH and DH). . . . . . . . . . . . . . . . 22" to 82" (55 cm – 208 cm)
Height Graduation (WB-300MH and DH) . . . . . . . . . . . . . . 1/4" (1.0 cm)
Power Source . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . (3) AA Alkaline batteries (included) for 

scale functions
(1) 3-volt CR123 Lithium battery (included) for
clock display

Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Dual panel LCD 1.0" display

(liquid crystal display)

Platform Size. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Extra Large 16" x 12" x 2-3/4" non-slip platform
Product Dimensions  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16.26" (413 mm) x 20.75" (527 mm) x 51.02" 

(1296 mm)

Product Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . WB-300

26.00 lb/11.80 kg

WB-300MH

27.50 lb/12.50 kg

WB-300DH

27.50 lb/12.50 kg

2. Important Notes for Users 

Handling Tips 

This scale is a precision instrument utilizing state-of-the art technology. To keep the unit in the best 
condition, follow these instructions carefully.

•  Do not attempt to disassemble the measuring platform and display.
•  Avoid excessive impact or vibration to the unit. 
•  Place the unit in an area free from direct sunlight, heating equipment, high humidity, or extreme 

temperature change. 

•  Never submerge in water or spill chemical liquids. To clean, use a dampened cloth and mild detergent.
•  If the product will not be used for a long period, remove the batteries before storage.

Moving the Scale

The scale has two rear wheels on the back of the platform. To avoid damage, use the wheels when moving
the scale. Hold the display head securely and tilt the scale back about 30-45 degrees. Wheels will allow
easy movement. IMPORTANT: Any time the scale is moved or re-located, be sure to press the ‘Zero’ key
before use.

Disabling Key Sounds

To confirm a key operation has been selected, a “beep” will sound when any key is pressed. To disable
this sound, press the up arrow and down arrow keys simultaneously for three seconds. To re-activate the
key sound, press the up arrow and down arrow keys simultaneously again for three seconds. The volume
of this beep sound is not adjustable.

Changing Measurement Units

This function can be used to change the measurement units on the Weight display: lb, kg, and fractional
lb. The measurement unit keys are located on the left display panel. You may press this key before or 
while the user is on the platform.

Press ‘LB’ to set to pound mode. Decimal weight value in pound mode (0.1 lb increments)
Press ‘KG’ to set to kilogram mode. Decimal weight value in kilogram mode (0.05 kg increments)
Press ‘LB (F)’ to set to fractional pound mode. Fractional weight value in pound mode (1/4 lb increments)

2

Summary of Contents for WB-300

Page 1: ...ortant Notes for Users 2 Handling Tips Moving the Scale Disabling Key Sounds Changing Measurement Units 3 Components 3 4 Key Operation 4 5 Assembly Instructions 5 6 Before Use 6 Level the Scale Insert the Batteries Remove the LCD Cover Set up the Clock Setting to Zero 7 Instructions for Use 8 Weighing Procedure Auto on Function Turn on Function Wake up Function Measuring Height Procedure 8 General...

Page 2: ... or extreme temperature change Never submerge in water or spill chemical liquids To clean use a dampened cloth and mild detergent If the product will not be used for a long period remove the batteries before storage Moving the Scale The scale has two rear wheels on the back of the platform To avoid damage use the wheels when moving the scale Hold the display head securely and tilt the scale back a...

Page 3: ...24 1 4 LB 2 3 4 1 WB 300 7 6 8 Standard Features WB 300 WB 300MH WB 300DH Measuring Platform Weight LCD Clock LCD Battery Compartment Height Rod Arm For WB 300MH and WB 300DH Height Rod Sliding Block Upper Pillar Column Four feet on the bottom of platform 1 2 3 4 5 6 7 8 WB 300MH WB 300DH 5 ...

Page 4: ...4 Key Operation 4 Weight LCD Right Side Clock LCD WB 300DH version only Right Side Clock LCD WB 300 and WB 300MH versions only ...

Page 5: ...illars as you bring them together Be careful not to pinch the coiled wire harness 8 between the two pillars 6 Assemble the height rod arm 7 to upper height rod 5 from the bottom of upper height rod 7 Assemble the top height rod assembly 5 to the lower height rod 4 Slide the assembled height rod into the height rod opening on the pillar assembly Push all the way down Note For scales without height ...

Page 6: ...ium battery in a separate bag The battery compartment is located on the back of the LCD display 1 Press down and pull battery compartment locking clip to remove the battery compartment cover 2 Insert batteries with the correct polarity 3 Replace the battery compartment cover and push it until you hear it click 6 AA alkaline AA alkaline AA alkaline LITHIUM BATTERY CB123A OR CR17335 SEE INSTRUCTION ...

Page 7: ...clock LCD Note After 5 seconds or more of keypad inactivity during this process the adjusted clock will be saved automatically When you change or re install the lithium battery the clock setting will be reset to 12 00 pm Please follow the instructions above to reset the clock Changing the 3 AA batteries will not cause the time setting loss Setting to Zero Zero function enables the user to reset th...

Page 8: ...l automatically shut off If the user remains on the platform the Weight LCD will display the weight for 20 seconds then automatically shut off Note If the weight is less than 30 lb the Auto On function will not work Please follow Turn on Function instructions listed below Turn on Function When the scale is in off mode and the user is not on the scale press any key to turn the power on 1 Press any ...

Page 9: ...LCD Hold sign will appear in the upper left corner of the Weight LCD 3 Step off the scale The Weight LCD will display the weight for 10 seconds then the scale will automatically shut off If the user remains on the platform the Weight LCD will display the weight for 20 seconds then automatically shut off Note DO NOT press any key while the scale is waking up and before the weight has been locked in...

Page 10: ...n the platform Place hand on the Height Rod Arm If the subject is taller than 44 pull the Height Rod Sliding Block and Height Rod Arm at the same time upward so that the arm is positioned higher than the subject s head If the subject is shorter than 44 move the Height Rod Arm down 10 INCORRECT way to hold the Height Rod Arm CORRECT way to hold the Height Rod Arm To avoid danger of breaking or comi...

Page 11: ...lar Column see example 1 If the subject is shorter than 44 read the number below the Height Rod Arm see example 2 11 INCORRECT way to measure the height of 44 or more CORRECT way to measure the height of 44 or more 92 36 35 91 90 89 88 144 57 58 44 43 145 146 147 112 111 110 109 108 Reading the height measurement of 44 or LESS Reading the height measurement of 44 or MORE Example 1 58 147 5 cm Exam...

Page 12: ...h clock for two seconds Please replace with three new AA alkaline batteries immediately Please refer to Inserting Batteries for instructions If you see Ad Er or nobAt on the weight LCD make sure the batteries are properly installed and have contact with the battery chamber The clock on this scale is operated by one replaceable 3 volt CR123 Lithium battery If the lithium battery is low LOWBATT icon...

Page 13: ...ale After battery replacement the scale may need to be reset in the following situation 1 If the batteries were installed in the incorrect polarity after being replaced in the correct polarity 2 When the low battery sign s is still displayed even after the new batteries are installed In order to reset the scale take a ball point pen to push a button hidden behind the lithium battery Please reprogr...

Page 14: ...Tanita France S A Villa Labrouste 68 Boulevard Bourdon 92200 Neuilly Sur Seine France Tel 33 0 1 55 24 99 99 Fax 33 0 1 55 24 98 68 www tanita fr Tanita Health Equipment H K Ltd Unit 301 303 3 F Wing On Plaza 62 Mody Road Tsimshatsui East Kowloon Hong Kong Tel 852 2838 7111 Fax 852 2838 8667 Tanita Europe GmbH Dresdener Strasse 25 D 71065 Sindelfingen Germany Tel 49 0 7031 6189 6 Fax 49 0 7031 618...