26

If problems occur, please refer to the following instructions before you ask for repairs.

 

Problem List

8.Troubleshooting

Problem

Solution

Nothing appears on the display after
pushing [

] key.

ON

OFF

• Please confirm that the Control Box is properly connected to the AC

Adapter and the AC Adapter is properly connected to the cable.

E-01 is displayed.

• E-01 is displayed when the impedance shows an abnormal value

compared to height and weight.
Please make sure that measurement was carried out with bare feet, the
electrodes or the soles of the feet were clean, and the feet were correctly in
contact with the electrodes. If the feet are dry or have calluses, place water on
the electrodes before stepping on the unit.

• It is possible that the body fat ratio exceeds the measuring capacity.

Please stop measuring.

E-02 is displayed and the reading of
the trunk cannot be obtained.

• This sign is displayed when the body fat percentage of the trunk is too

low and is off the measureable range.

• Please refer to the readings of other parts for the body fat.

27

RS-232C Interface Instructions

This section describes the interface for outputting the BC-418 Body Composition Analyzer measurement results as
RS-232C signals to personal computers and other external devices.

CAUTION

Only connect IEC950-compliant computers to the BC-418.

•Connection example 

Always use a reverse cable when connecting to an external computer.

 

Specifications

 

Signal lines and connection methods

Communication Standard

EIA RS-232C Compatible

Asynchronous Transmission

4800 baud

8 bits

None

1 bit

CR+LF

Communication Method

Signal speed

Data bit length

Parity

Stop bit

Terminator

Terminal number

Signal name

2

RXD

3

TXD

5

GND

1.   

RXD
TXD

GND

2.   RXD
3.   TXD
4.  
5.   GND

Personal computer or
other device

•An RS-232C connector (D sub 9-pin male) is located in the lower center on the back of the control box.
•Use an RS-232C cross cable when connecting to a personal computer or other device.

D sub 9-pin female

RS-232C cross cable

RS-232C connector (D sub 9-pin male)

E-11 is displayed.

• E - 11 is displayed when there is a loose connection between the control

box and the weighing platform.

• Confirm that none of connections between the scale and control box are

loose or unplugged.

• There may be excessive vibration which will disturb the measurement process.

E-12, 13 or 14 is displayed.

• The equipment requires readjustment.

E-16 is displayed.

• Measurement is not possible because impedance is unstable. In order to

improve electrical conductivity, please clean the soles of the feet or
place water on the electrodes before using the equipment.

No print out

• Please confirm the number of print outs selected is more than 0. (See P. 13)
• Please confirm that the correct brand of paper is used.
• Please confirm that the printer paper roll is placed in the correct position.
• Please confirm that the printer paper is not jammed.

The printer paper does not come out.

• Please confirm the printer paper is not jammed.

P.End is displayed and the equipment
is incapable of taking a measurement.

• The printer paper is used up. Please put in a new printer paper roll. (See P.11)
• Please confirm that the printer paper is being advanced properly.
• Please make sure that the Paper Release Lever is not in the “Up” position. (See P. 25

)

---- is displayed.

• It is possible that the maximum weight capacity has been exceeded.

Please stop measuring.

uuuu is displayed.
The power has turned off during measurement.

• Do not stand on or put things on the Weighting Platform before starting measurement.

Start measurement after confirming there is nothing on the Weighing Platform.

Feed Key is not functioning.

• It is possible that the number of print outs selected is 0 or the Weight

Scale Mode has been selected. Please select the Body Fat Measurement
Mode and select a number larger than 0 for the number of print outs.

Summary of Contents for BC-41

Page 1: ... Please read this Instruction Manual carefully and keep it handy for future reference BODY COMPOSITION ANALYZER BC 418 INSTRUCTION MANUAL ...

Page 2: ... Only Function 21 Explanation of the Print Out 22 7 Clearing Paper Jams 24 Components of the Printer Unit 24 Clearing Paper Jams 25 8 Troubleshooting 26 Problem List 26 RS 232C Interface Instructions 27 Technical Notes 36 THE APPLICABILITY OF THE BMR REGRESSION FORMULA AND ITS DIFFERENCES WITH THE OLD FORMULA 41 Product specifications 43 Advantages 1 BIA is a simple quick and non invasive means of...

Page 3: ...ons for Accurate Measurement This equipment sends out a very weak electric current to measure impedance electrical resistance of the body Therefore in principle users need to use this equipment with bare feet Moreover since impedance fluctuates in accordance with the distribution of body fluid please observe the following instructions for accurate measurement To prevent a possible discrepancy in m...

Page 4: ...cm ft in AGE FAT lb Body weight mark Indicates that the device is being used to measure body weight Clothes weight mark Indicates that the weight of clothes is being subtracted Gender mark male Indicates that the user is male Gender mark female Indicates that the user is female Athlete mark Indicates that the user is an athlete Ready mark Indicates that the device is ready to begin measurements Me...

Page 5: ...tly exercise less than 10 hours per week Tanita s athlete definition does not include enthusiastic beginners who are making a real commitment to exercising at least 10 hours per week but whose bodies have not yet changed to require the Athlete mode Please see Technical Notes booklet for further explanation Numerical Keys Inputs numerical values Time Set Key Sets the date and time CLOTHES ATHLETIC ...

Page 6: ...uge Check DC Jack for AC Adapter Adjustable feet 10 Loading the Printer Paper Roll CAUTION Please change the paper roll when red lines appear along the sides of the paper FEMALE MALE ATHLETIC STEP ON AGE FAT kg lb ft in cm CLOTHES WEIGHT 3cm 1 Turn the equipment on by pressing the key P End will flash in the middle of the display This means that there is no printer paper When you do not have any p...

Page 7: ...the printout language and the printout items AGE FAT ATHLETIC STEP ON FEMALE MALE kg lb ft in cm CLOTHES WEIGHT AGE FAT ATHLETIC STEP ON FEMALE MALE kg lb ft in cm CLOTHES WEIGHT AGE FAT ATHLETIC STEP ON FEMALE MALE kg lb ft in cm CLOTHES WEIGHT AGE FAT ATHLETIC STEP ON FEMALE MALE kg lb ft in cm CLOTHES WEIGHT AGE FAT ATHLETIC STEP ON FEMALE MALE kg lb ft in cm CLOTHES WEIGHT AGE FAT ATHLETIC STE...

Page 8: ...MALE MALE kg lb ft in cm CLOTHES WEIGHT H I N T The target body fat ratio function is deactivated when this product is shipped from factory Setting the Original Mode 1 Press the key while holding down the key 0 The target body fat ratio function is deactivated 1 The target body fat ratio function is activated If the number of print outs is set to 0 during the process of Setting the Number of Print...

Page 9: ...t Fat Mass FFM Predicted MuscleMass Left Arm Fat Fat Mass FFM Predicted MuscleMass Trunk Fat Fat Mass FFM Predicted MuscleMass 551 Ω 212 Ω 214 Ω 292 Ω 309 Ω 8 9 1 1 kg 11 5 kg 10 9 kg 10 1 1 2 kg 11 0 kg 10 4 kg 14 0 0 6 kg 3 6 kg 3 4 kg 15 4 0 6 kg 3 5 kg 3 3 kg 15 0 6 0 kg 34 1 kg 32 8 kg 20 79 6 kg 15 9 kg 6 3 kg TANITA BODY COMPOSITION ANALYZER B C 4 1 8 C o n s u l t y o u r p h y s i c i a n...

Page 10: ...If the number of print outs is set to 0 nothing will be displayed If the target body fat ratio is set to OFF the target body fat ratio will not be printed out Before you start a body weight management program and set the appropriate personal body fat ratio please consult your doctor Tanita is not responsible for setting the appropriate target body fat ratio for specific individuals For details of ...

Page 11: ...s of the body fat ratio or the quantity of fat are abnormally small or the error message E01 is shown on the display the probable reason is that the soles of the feet and the electrodes are not in full contact Make sure you step on the Weighing Platform so there is contact between the electrodes and the soles of your feet If the problem is not solved this way it is possible that the soles of the f...

Page 12: ... is calculated by the following formula Weight kg Height m2 Desirable Range 18 5 24 9 IMPEDANCE Impedance reflects the body s inherent resistance to an electrical current Muscle acts as a conductor of the electrical current adipose tissue acts as a resistor BMR Basal Metabolic Rate represents the total energy expended by the body to maintain normal functions at rest such as respiration and circula...

Page 13: ...ever the Automatic Cutter Unit and the Printer Cover to the normal position Insert the catches of the Printer Cover into the guide holes of the Control Box and push it down slowly while pressing the catches on both the left and right hand sides until a click is heard Failure to press down the Paper Release Lever will result in continuous feeding of the printer paper without it being cut automatica...

Page 14: ...ated in the lower center on the back of the control box Use an RS 232C cross cable when connecting to a personal computer or other device D sub 9 pin female RS 232C cross cable RS 232C connector D sub 9 pin male E 11 is displayed E 11 is displayed when there is a loose connection between the control box and the weighing platform Confirm that none of connections between the scale and control box ar...

Page 15: ...hlete 1 male 2 female 1 male 2 female Age Output data ASCII code 28 2 Output data format Measurement data will be output in the following format A comma is used as a separator between data items The terminators marking the end of data are CR ASCII code ODH and LF ASCII code OAH If the digits in the data do not reach the stipulated length for fixed length data the data will be right justified for o...

Page 16: ...tempted to start measuring without completing settings E01 Error in calculating body fat percentage EXX other error body fat analyzer switches off LF CR X X E LF CR 1 0 E LF CR 0 0 E See p 26 7 Troubleshooting for details of errors 31 Output Fat Monitor Mode Format Remote Mode All types of data may be entered by connecting the BC 418 to a PC and switching to Remote mode Switching to Remote Mode 1 ...

Page 17: ...3 4 D Host body composition analyzer LF CR 8 1 6 D Host if data settings incorrect body composition analyzer If settings incorrect Ex 18 Host body composition analyzer LF CR 8 1 6 D Host body composition analyzer LF CR 6 D Host body composition analyzer LF CR 6 D If commands accepted If settings incorrect Set target BF 8 33 Switch on body composition analyzer 1 Choose the units of measurement 2 Fi...

Page 18: ...e body composition analyzer During measurement 11 Measurement finished 12 After measurement has finished all necessary items of data and tare weight are cleared and the unit waits for further commands body composition analyzer 2 1 0 9 0 2 1 9 2 9 0 1 3 4 0 0 1 7 9 Date Time Height Age 0 7 3 3 1 3 1 0 0 9 6 0 6 3 5 5 1 2 1 2 2 1 4 2 9 2 3 0 9 7 0 4 6 6 Weight Fat mass Fat free mass Body water mass ...

Page 19: ...s supplied from the tips of the toes of both feet and the fingertips of both hands and the voltage is measured on the heel of both feet and the thenar side of both hands This method allows five different impedance measurements to be made whole body right leg left leg right arm and left arm by switching the part of the body in which the current is flowing and the location where the voltage is measu...

Page 20: ...those particular body parts The reasons for changes in impedance are different for the upper limbs and the lower limbs Given that the impedance of the trunk is extremely low a mere 5 10 of the impedance between hands and feet changes involving the trunk can be almost wholly ignored when measuring the impedance between hands and feet and the changes for the upper and lower limbs are synthesized Whe...

Page 21: ...tra day fluctuations FAT 22 21 20 19 18 17 10 00 18 00 16 00 14 00 12 00 5 13 Monday 5 14 Tuesday 5 15 Wednesday 5 16 Thursday Inter day changes in body fat percentage and impedance measured by BIA between the feet during normal daily routine and during dehydration 40 THE APPLICABILITY OF THE BMR REGRESSION FORMULA AND ITS DIFFERENCES WITH THE OLD FORMULA The New Regression Formula for Basal Metab...

Page 22: ...ows 5 Digits LCD Output Data Interface RS 232C D sub 9 pins male Connector Temperature Range of Usage 0ºC 35ºC Power Consumption Relative Humidity 17 5 W BC 418 Power source AC adaptor included Center Minus MODEL SA25 0535U CLASS 2 Input Voltage 100 240 VAC 50 60 Hz 550mA Output Voltage 5 VDC 0 3 V Rated Current 3 5 A No Load Input Voltage 5 VDC 0 3 V 30 80 without condensation Weight of Equipment...

Page 23: ...189 71 Unit 301 303 3 F Wing On Plaza 62 Mody Road Tsimshatsui East Kowloon Hong Kong Tel 852 2838 7111 Fax 852 2838 8667 2625 South Clearbrook Drive Arlington Heights Illinois 60005 U S A Toll Free 800 826 4828 Tel 847 640 9241 Fax 847 640 9261 http www tanita com Villa Labrouste 68 Boulevard Bourdon 92200 Neuilly Sur Seine France Tel 01 55 24 99 99 Fax 01 55 24 98 68 The Barn Philpots Close Yiew...

Reviews: