14
1.6
Metering
The KV mk2 series range use a tap on the secondary of the HV transformer for voltage
metering. This can be more accurate than primary metering, although capacitive loads will still
cause the meter to read low.
If very accurate kV metering is required, it is necessary to use an external HV divider.
All of the units use moving coil analogue meters.
1.6.1
Metering Ranges
Unit
Voltmeter FSD
mA meter FSD
X1
X0.5
X1
X0.1
KV30-100
40kV
20kV
120mA
12mA
KV50-100
60kV
30kV
120mA
12mA
KV50-200
60kV
30kV
240mA
24mA
KV100-100
120kV
60kV
120mA
12mA
1.6.2
IMPORTANT NOTE - Metering Connections
It is essential that the metering lead from the high voltage transformer to the control unit is
connected in circuit. If this lead is not connected, the metering on the control unit will not
function and the variable overload trip will not function.
Summary of Contents for KV mk2 Series