background image

INDICATIONS FOR USE

The AutoSoft

 90 infusion set is indicated for the subcutaneous infusion of insulin administered by an external pump.

DESCRIPTION

AutoSoft 90 is an infusion device which combines an infusion set (catheter) with an aid for insertion (inserter). As the infusion 

set is an integrated part of the inserter, AutoSoft 90 is delivered assembled and ready for use as a single, sterile unit.

CONTRAINDICATIONS

AutoSoft 90 is neither intended nor indicated for use in intravenous infusion of blood or blood products.

WARNINGS

•  AutoSoft 90 infusion sets with t:lock tubing connectors should only be used with Tandem cartridges featuring the t:lock 

connector.

•  AutoSoft 90 is only sterile and non-pyrogenic if the insertion device is unopened and undamaged. Do not use if the  

insertion device is already opened or has been damaged. Ensure sterility by checking if the sterile paper and tamper-proof  

seal are undamaged.

•  Carefully examine the instructions for use before inserting AutoSoft 90. Failure to follow instructions may result in  

pain or injury.

•  When using AutoSoft 90 for the first time, do so in the presence of a healthcare provider. Since different individuals have 

varying amounts of subcutaneous layer, ensure that an appropriate cannula length is chosen to fit your needs.

•  Do not in any way put disinfectants, perfumes, deodorants, cosmetics or other products containing alcohol or disinfectants 

in contact with the connector or the tubing. These may affect the integrity of the infusion set.

•  AutoSoft 90 is a single-use device and should be disposed of immediately after use. Do not clean or re-sterilize.
•  AutoSoft 90 can be disposed of safely after insertion if the lid has been placed back on. Please consult your local pharmacy 

for sharps containers.

•  Be sure that the needle guard is removed before insertion is made.
•  Release the tubing with caution as a hard pull of the tubing can result in damage to the infusion set/introducer needle. 

Ensure that the infusion set is properly in place when the tubing is fully released 

(see figure 8)

.

•  Do not leave air in the infusion set. Make sure to fill AutoSoft 90 completely.
•  Inaccurate medication delivery, infection and/or site irritation may result from improper insertion or maintenance of  

your infusion site.

•  Replace the infusion set every 48-72 hours, or per your healthcare provider’s instructions.
•  If the soft cannula bends during insertion, discard and apply a new AutoSoft 90 immediately.
•  Replace the infusion set if the adhesive tape becomes loose or is displaced from its original site. Since the cannula is soft, 

it will not cause any pain if it slips out, and this may take place without notice. The soft cannula must always be completely 

inserted to receive the full amount of medication.

•  If your infusion site becomes inflamed, replace infusion set and use new site until the first site has healed.
•  Do not re-insert the introducer needle into the infusion set. This could cause a tearing of the soft cannula and unpredictable 

medication flow.

•  Never try to fill or free clogged tubing while the infusion set is inserted. This may result in unpredictable medication flow.
•  Always clean the insertion site when temporarily disconnecting the infusion set. Consult your healthcare provider on how  

to compensate for missed medication when disconnected. Carefully monitor blood glucose levels when disconnected  

and after reconnecting.

•  Protect the AutoSoft 90 infusion set from direct sunlight and atmospheric humidity. Store in a dry place at room temperature.
•  Never point a loaded insertion device towards any body part, where insertion is not desired.
•  Reuse of the infusion set may cause infection, site irritation, or damage to the cannula/needle. A damaged cannula/needle 

may lead to inaccurate medication delivery.

RECOMMENDATIONS

•  Check your blood glucose level 1-2 hours after introducing AutoSoft 90. Measure your blood glucose on a regular basis. 

Discuss this with your healthcare provider.

•  When using insulin, do not change your infusion set just prior to bedtime, unless blood glucose can be checked 1-2 hours  

after insertion.

•  When you are infusing insulin and your blood glucose level becomes unexplainably high or an occlusion alarm occurs, check 

for clogs and leaks. If in doubt, change your infusion set since the soft cannula could be dislodged, crimped or partially 

clogged. Discuss a plan for rapid replacement of insulin with your healthcare provider should any of these problems arise. 

Test your blood glucose level to make sure the problem is corrected.

•  Carefully monitor your blood glucose level when disconnected and after reconnecting.

PREPARATION AND INSERTION

Read all instructions carefully before using the AutoSoft 

90 infusion set. For information about pump therapy, 

possible errors, and potential risks and their causes, 

consult your pump manufacturer’s instructions for use 

before connecting AutoSoft 90. When using AutoSoft 

90‚ follow proper hygiene procedures. Please see the 

corresponding illustrations.

INSTRUCTIONS

1. 

Wash your hands before inserting AutoSoft 90.

2. 

Recommended insertion areas. Do not use an area 

right next to your previous place of insertion.

3. 

Clean the area of insertion with a disinfectant as 

directed by your healthcare provider. Be sure the 

area is dry before inserting AutoSoft 90.

4. 

Pull the red tape to remove the seal.

5. 

Remove the sterile paper.

6. 

Press the three raised dots on each side of the lid 

firmly with one hand and lift up with the other hand.

CAUTION:

 Be careful not to bend or touch the 

AutoSoft 90 introducer needle when preparing the 

inserter and infusion set.

IMPORTANT NOTE:

 If the infusion set is not 

securely placed in the insertion device with the 

needle pointing straight ahead prior to insertion, 

pain or minor injury may occur. To avoid this, 

please ensure that the following steps are taken 

gently so the infusion set is not loosened / 

removed by accident.

7.  UNWINDING THE TUBING:

 Carefully twist the 

beginning of the tubing out of the slot. Pull up 

gently. Now unwind the tubing from the insertion 

device by pulling it carefully upwards.

CAUTION:

 Do not touch the AutoSoft 90 

introducer needle when unwinding the tubing. 

Be careful not to pull too hard when the tubing is 

fully released. You could accidentally remove the 

infusion set from the introducer needle.

8. 

Connect AutoSoft 90 to a reservoir/cartridge and 

the pump. Prime infusion set. Ensure there are no 

air bubbles in the tubing or AutoSoft 90.

CAUTION:

 When priming the infusion set, hold 

AutoSoft 90 with the needle pointing down to 

ensure that medication does not come in contact 

with the adhesive backing paper.

9. 

Pull up gently to remove the adhesive backing 

paper.

10.  PREPARE THE INSERTER:

 Place fingers on lined 

indentations. Press lined indentations on each side. 

Pull spring up until you hear a “click”.

11. 

Carefully remove the needle guard by gently 

twisting and then pulling it.

CAUTION:

 Before positioning AutoSoft 90 over 

clean insertion site, place the tubing in the slot to 

ensure that the tubing is not caught underneath 

the device during insertion.

12.  INSERT AUTOSOFT 90:

 Position AutoSoft 90 over 

the insertion site. Simultaneously press the round 

indentations on both sides of the insertion device 

to insert AutoSoft 90.

13. 

Push gently on the center of the inserter to secure 

the adhesive on the skin.

14. 

Remove the inserter/introducer needle by grasping 

the center and pulling it gently back. Massage the 

tape thoroughly onto the skin. Fill the empty space 

in the cannula with the following amount of insulin:

15. 

Push the lid back in place until you hear a “click.”

U-100 insulin:

6 mm: 0.3 units (0.003 ml)
9 mm: 0.5 units (0.005 ml)

DISCONNECT

A. 

AutoSoft 90 allows you to temporarily disconnect your pump without having to 

change the infusion set. Hold the adhesive in place by placing a finger just in front of 

the cannula housing, and gently squeeze the sides of AutoSoft 90. Pull the connector 

needle straight out from the cannula housing.

B. 

Insert the disconnect cover into the cannula housing until you hear a “click”.

RECONNECT

C. 

Prime the infusion set until medication dispenses from the needle.

CAUTION:

 When priming the infusion set, hold AutoSoft 90 with the needle 

pointing down.

D. 

Place a finger on the adhesive tape just in front of the cannula housing.

E. 

Push the connector needle straight in until you hear a “click”.

Use by (year-month-day)

Do not re-use

Sterilized using ethylene oxide

Attention: See accompanying documents

See Instructions for Use

Reference / model number

Lot number

Manufacturer

© 2017 Tandem Diabetes Care, Inc. All rights reserved. Tandem Diabetes Care is a registered 
trademark and AutoSoft is a trademark of Tandem Diabetes Care, Inc. AW-1000184_A

Unomedical a/s

Aaholmvej 1-3, Osted  

4320 Lejre, Denmark
Made in Mexico

Legal manufacturer:

Distributed by:

11045 Roselle Street 

San Diego, CA 92121, USA

www.tandemdiabetes.com 
+1-858-255-6269

Instructions for Use

CONTENTS

A. 

Lid

B. 

Needle guard

C. 

Soft cannula

D. 

Adhesive

E. 

Cannula housing

F. 

Connector needle

G. 

Tubing

H. 

Introducer needle

I. 

Insertion device

J. 

Sterile paper

K. 

Three raised dots – press here  

when lifting the lid off 

(see figure 6)

L. 

Lined indentations – press here  

when loading AutoSoft 90  

(see figure 10)

M. 

Round indentations – press here  

when inserting AutoSoft 90  

(see figure 12)

N. 

Disconnect cover  

(see figure B)

Reviews: