T&D TR-81 Instruction Manual Download Page 6

10

RECORDING

The REC indicator comes on when recording is begun. It blinks while on
standby. The amount indicator marks increase as the data increases.

9

Recording Indicator.

Amount of Recorded Data.

Explanation of the LCD Display

DISPLAY

COMMUNICATING

The COM mark appears when communication is taking place between the
unit and your computer, i.e. when setting up recording, or downloading
data.

BATTERY LIFE WARNING

The BAT message appears whenever the battery power is low. Record-
ing is still possible while lamp is on, but when blinking recording will be
discontinued. Please change the battery as soon as you see the warn-
ing lamp come on.

FULL

The FULL mark appears when 8000 units of data have been recorded
in ONETIME Mode. This indicates that recording has stopped and it is
important to download the data into your computer as soon as possible.

OVER THE MEASURABLE RANGE

The Hi mark indicates that the temperature is over the measurable range
of 600

.

BELOW THE MEASURABLE RANGE

The Lo mark indicates that the temperature is below the measurable
range of -200

.

SLEEP MODE

The battery lamp will begin to blink when power is very low, recording
will stop and the unit will automatically go into sleep mode. It is impor-
tant to change the battery before it begins to blink.

Indicates communication in progress.

Battery Life Warning.

Displays Data Measurement for Channel 1.

Displays Data Measurement for Channel 2.

Temperature Unit Display.

Display unit can be changed via. computer.

Sensor Type  (Pt100)

Recording Mode

ENDLESS: When the recorded data goes over 8000 units the oldest data is

overwritten and recording continues.

ONETIME: When the recorded data reaches 8000 units the REC indicator goes off

and recording stops.

One Time  Mode

Changes to 

FULL and recording

stops.

Endless Mode

  Indicates recording will continue.

Summary of Contents for TR-81

Page 1: ...1999 12 16004144020 Thank you for buying our product Please read carefully before using Copyright 1999 T D Corporation All rights reserved Instruction Manual...

Page 2: ...erference received including interference that may cause undesired operation NOTE This equipment has been tested and found to comply with the limits for a Class A Digital Device pursuant to Part 15 of...

Page 3: ...s into the unit may cause fire or malfunctioning If something does leak into the unit make sure to remove the batteries immediately Maintenance or modification of the main unit by anyone other than ou...

Page 4: ...commercial redistribution should be directed to the Sales Department of T and D Corporation Please do not attempt to make any changes or modifications to Hi for Windows This product is a thermo recor...

Page 5: ...ws 1 Lithium Battery 1 Attachment Plate 1 screw Contents of Package CR2 1 Instruction Manual Guarantee Use the attachment plate to hang the Thermo Recorder on the wall etc How to use the Attachment Pl...

Page 6: ...ata into your computer as soon as possible OVER THE MEASURABLE RANGE The Hi mark indicates that the temperature is over the measurable range of 600 BELOW THE MEASURABLE RANGE The Lo mark indicates tha...

Page 7: ...re represented with the letters A B or color red white After connecting the sensor tighten the screws to prevent slipping 11 Install the Battery Preparing the Thermo Recorder Using a coin turn the bat...

Page 8: ...ows 3 1 95 98 NT 3 51 NT 4 0 is not activated properly it may be impossible to install and run Hi for Windows Quit all other applications If you are running other applications make sure to quit them b...

Page 9: ...g the gauge you can move up and down to the data you want to be displayed Magnification and Widening Using the Mouse Magnification With the left button drag the mouse to outline the area you want to m...

Page 10: ...properly pushing in until you hear a click 17 If you connect to the wrong place communication will not take place Also make sure that the cable is connected properly in the right position Side view of...

Page 11: ...seconds 10 minutes If you press on the REC STOP button when the REC lamp is not lit the next recording will begin and any former data will be automatically deleted Start Recording via Base Unit When...

Page 12: ...ng temperature and others This explanation is based on using a new battery and in no way should be taken as a guarantee of battery life Recording Int of more than 2 sec s Recording Int of 1 second Max...

Page 13: ...e if you can download using it A If you have a computer with energy saving function settings make sure that the serial port has not been turned off Especially on NEC brand PC98 notebook computers the...

Page 14: ...450 when main unit is at 20 10 1 0 450 600 No account for sensor differences related to sensor type Measurement Resolution 0 1 Measurement Current 1 mA Starting Methods Immediate Programmed by compute...