Temperature Sensor Replacement Procedures and Cautions
1.
Fit the supplied Sensor Replacement Tool to the bottom of the Sensor. Then turn it in the
counterclockwise direction to loosen the Sensor from the Main Unit. Remove the Sensor from
the Main Unit slowly.
Rubber Packing
Sensor Spacer
Cord (Red / White)
Turn in the counterclockwise
direction
2.
Pull out the Connector from the inside of the RTR-61 Main Unit.
Connector
-
DO NOT forcibly pull the cord. Doing so may break it
off.
3.
Hold the connector securely to disconnect. DO NOT pull the cord.
4.
Change the Sensor Spacer and the Rubber Packing, if necessary. Connect the cord of the
replacement sensor to the connector.
- To avoid poor connections, make sure to the
Connector is securely connected.
- Make sure that the cord is properly connected by
inserting it until you hear a clicking sound.
5.
Insert the cord into the RTR-61 Main Unit as shown in the fi gure.
Insert it into the RTR-61 Main
Unit as it is.
How to prevent the cord from breaking due to kink
Straighten the cord. Then twist the cord two
and a half times in the counterclockwise
direction, and
6.
Turn the bottom of the Sensor slowly in the clockwise direction. At the end, be sure to fasten it
tightly using the supplied Replacement Tool.
Caution: DO NOT Cause Sensor Spacer Misalignment!
Align the Sensor Spacer with the shape of the RTR-61 Unit.
Make sure that there is NO space between the Sensor Spacer
and the RTR-61 Unit.
- Misalignment of the Sensor Spacer and looseness of
the Sensor bottom will allow water or foreign objects
to enter into the unit.
If water or a foreign object enters into the unit, immediately
remove the batteries and stop using it.Continued use may cause
fi re, electrocution or damage.