10. Appendix

Fuse Replacement on Maelstrom 9610

Specification of Fuse:

UL recognized component

Rated 10A, 250VAC x2pcs

UFE, Size 5.2

Φ

x 20mm

Precautions of Fuse Replacement

Please note that severe electrical shock which may cause physical 
injury or death could be resulted from working on an instrument as 
the high voltage power supply is in operation. To avoid electrical 
shock, please follow the guide below to replace the fuses:
1. Disconnect the power to the instrument
2. Unplug the power cord from the instrument
3. Still the instrument at least 1 min before fuse replacement

Procedure of Fuse Replacement

Required tool: Fine flat-tip screwdriver
1. Power off and unplug the power cord from the instrument
2. Still the instrument for at least 1 min
3. Use flat-tip screwdriver to slot in power filter module and pull 

4. outward the fuse holderPush the fuse aside and replace the brown 

fuse by new one

5. Place the fuse compartment back to the filter module till it locks in 

place

Fuse

Fuse

ENG

LI

SH

Summary of Contents for Maelstrom 9610

Page 1: ...Maelstrom 9610 User Manual TAN Bead Nucleic Acid Extractor Non Sterile 039 M1602 X01Rev V1 0_2021 06 11 Compatible with power 100V 120V only...

Page 2: ...ent Overview 11 3 Installation Get Started 15 4 On screen Display How to Use 16 5 Technical Support 38 6 Cleaning and Maintenance 39 7 Disposal 39 8 Patents 40 9 About Manufacturer 40 10 Appendix 41 2...

Page 3: ...ty and environmental protection standards for products sold within the European Economic Area EEA In Vitro Diagnostic Medical Device Consult Instructions for Use Authorized Representative in the Europ...

Page 4: ...the power cord Before working on maintenance must disconnect the power plug from the outlet in advance Do not pour any liquid on the instrument Do not place any containers with liquid on the instrumen...

Page 5: ...so may cause malfunctions Repairs to the instrument must be performed by agencies authorized by Taiwan Advanced Nanotech Inc Only original spare parts from Taiwan Advanced Nanotech Inc can be used on...

Page 6: ...c rods to collect and transfer magnetic beads and spin tips to mix suspension Rotating the spin tips will mix suspension very well Extended magnetic rods can collect magnetic beads efficiently Purifie...

Page 7: ...culation Place Maelstrom 9610 on the table capable of bearing at least the weight of Maelstrom 9610 95 kg Do not use Maelstrom 9610 in a location where there is large temperature variability or high h...

Page 8: ...should adhere to their institutional guidelines for the handling and disposal of all infectious substances used with this device It is important to clean the device after every use If samples or reag...

Page 9: ...for electrical equipment for measurement control and laboratory use Part 2 101 Particular requirements for in vitro diagnostic IVD medical equipment EMC Requirements The medical device has passed the...

Page 10: ...om region to region Equipment list 10 1 Introduction Power cord x1 Remark Power cord varies from region to region Auto Tube base is required in use of Auto Tube reagent kits Tips combo pack x1 Magneti...

Page 11: ...I Fuse 250V 10A Max Throughput 96 samples per run Process volume 50 l 1600 l Spin speed up to 3000 rpm Heater 4 independent heating plates Magnetic rod 3900 gauss Display 7 inch touchscreen UV HEPA A...

Page 12: ...urntable contains 8 plate holders 4 plate holders 1 2 3 6 have the heating module for temperature control The plate holder no 8 is for mounting and ejecting spin tips Power Button USB I O USB type A p...

Page 13: ...pe A for use of upgrading HMI program USB type A for use of upgrading PLC program Power inlet Fuse Please note that only TANBead technicians and authorized technicians or service engineers shall engag...

Page 14: ...o initiate the program To turn off the alarm To pause the program Press again to resume the program To turn on off the LED light Emergency to stop machine 14 2 Instrument Overview ENGLISH Control Pane...

Page 15: ...le is highly advised Step 2 Remove the cushions around the instrument Step 3 Place the instrument to a flat table that can bear 120kg weight Step 4 Connect the power Please note that this instrument i...

Page 16: ...and new machine there is a need to create an user in advance Click the icon to create your own user ID Tab the icon to create a new user ID Click blank space to call out keyboard Use the keyboard to i...

Page 17: ...program will automatically lead you access the system Instrument is trying to save new user ID After save you will be guided to the main page Now you can prepare for extraction work Switch user To sw...

Page 18: ...ser ID not found Make sure you input correct existing user ID account 4 On Screen Display and How to Use User ID repeated Please change the name of creating an ID Account is full Max 24 user ID are al...

Page 19: ...d ejecting spin tips To execute the extraction program To export extraction records To use UV HEPA to sterilize the instrument To edit add and delete extraction program To set basic parameter of instr...

Page 20: ...int Setting Parameter setting value upon ejecting tips Eject Tip Point Actual Actual tip ejection parameter 4 On Screen Display and How to Use Tip function is designed to test the performance of mount...

Page 21: ...e underneath the extraction module Step 4 Click Mount icon to execute mounting spin tips tip combo pack required or click Eject icon to execute ejecting spin tips blank 96DW plate required Tip functio...

Page 22: ...lstrom 9610 is capable of storing up to 25 programs Please follow the guideline of reagent kits and correctly select the program Step 2 Click Next to continue 4 On Screen Display and How to Use Run fu...

Page 23: ...Tab the circle first and then use on screen keyboard or USB barcode scanner to scan in if 1 D barcode is available Sample ID can bear 18 characters input Once completed tab Next to continue Remark Ma...

Page 24: ...be placed is always the tip combo pack Step 6 Ready to run the extraction program Click Yes to begin the operation If you choose No the process will go back to previous page 4 On Screen Display and H...

Page 25: ...e of extraction program started Est End time Estimated time the extraction program coming to the end Speed Spin speed Volume Liquid volume in the working well Pause Cease an extraction program tempora...

Page 26: ...n the instrument will not act immediately until the collection process comes to the end Vaporing EtOH Before entering Elute process the protocol will must work on Vapor process to evaporate EtOH Poor...

Page 27: ...ed to store the last 15 experiment records By clicking View button you can navigate the detail of sample ID Report function enables you to view experiment records and the identity that performed the e...

Page 28: ...lighting Before turn on UV please must close the cover door of turntable The recommended lighting time is at least 30min The more the better It is highly advised to use UV to sterilize the instrument...

Page 29: ...traction steps time of mixing and collect To add a new program you need to use Copy to duplicate a program first and then use Edit to write the protocol Delete a program no longer used to save the mem...

Page 30: ...g process S Spin SS Spin switch Spin direction will change regularly S U D Spin with up and down move SS U D Spin switch with up and down move Name Name of plate which following buffer may be involved...

Page 31: ...t plate is about to be used Mixing Time of mixing range from 0 to 999 9 mins Spin Speed of spin mixing range from 500 to 3000 rpm Collect Time of collecting beads range from 0 to 999 9 sec Vapor Time...

Page 32: ...umber to add the program You still can tab destination number to alter The keyboard will appear after tab on destination number Once complete click Yes to add the program 1 2 Please find a number to a...

Page 33: ...33 Delete a Program To delete a program tab the program no longer need and then click Delete icon to execute Click Yes to confirm the deletion 1 2 ENGLISH 4 On Screen Display and How to Use...

Page 34: ...anner management 4 On Screen Display and How to Use Setting Language Maelstrom 9610 supports multi language display including English Traditional Chinese Simple Chinese Korean and Spanish Time When a...

Page 35: ...tch display language 2 1 Tab Language icon to access ENGLISH 4 On Screen Display and How to Use Setting Language Maelstrom 9610 supports multi language operating interface including English Traditiona...

Page 36: ...ime Set correct time for the system according to the local time Click icon to set Date and Time for the system once set up is done tab OK to apply the latest setting 1 2 4 On Screen Display and How to...

Page 37: ...with USB type infrared barcode scanner Click the icon to ON OFF of barcode scanner If barcode scanner is set OFF even USB type barcode scanner is connected to Maelstrom 9610 scanned data will not be p...

Page 38: ...r our company for more help If there is a need to contact us you can call our rep number at 886 3 3167568 or email us at service tanbead com for further assistance To efficiently solve the problems an...

Page 39: ...rods unprotected If this happens the magnetic rod needs to be cleaned immediately To clean the magnetic rods wipe with a soft cloth dampened with pure water Do not use alcohol solvent If the magnetic...

Page 40: ...49 511 3908 9530 9 About Manufacturer ENGLISH Patent List USA US09616398B2 EU EP2937136 Canada CA2862946 Japan JP6151735B2 Korea KR101696517B1 China CN104971638B Taiwan TWI526245B WIPO WO2016127292 8...

Page 41: ...uide below to replace the fuses 1 Disconnect the power to the instrument 2 Unplug the power cord from the instrument 3 Still the instrument at least 1 min before fuse replacement Procedure of Fuse Rep...

Page 42: ...ind the folder named Maelstrom Master Then double click the utility TANBead Maelstrom 9610 Master Connect the instrument to your PC Power on the instrument and then use the cable USB A to B to link th...

Page 43: ...ment Program Editing After connection the utility program will automatically read all protocols stored in the instrument Select the protocol first and then choose an editing function to apply Please n...

Page 44: ...he instrument Remember to SAVE settings After adjustment of protocol contents please MUST click Save to get most updated contents effective Once you click Save the utility program will automatically u...

Page 45: ...Maelstrom 9610 100 120V 039 M1602 X01 Rev V1 0_2021 06 11...

Page 46: ...1 50 2 55 3 59 4 60 5 82 6 82 7 83 8 83 9 84 10 85 46...

Page 47: ...47 EEA WEEE WEEE...

Page 48: ...AC 48...

Page 49: ...49...

Page 50: ...Maelstrom 9610 DNA RNA 50 1...

Page 51: ...1 120 10 35 40 80 5 60 20 90 2000 51...

Page 52: ...80 52 1...

Page 53: ...IEC 61010 1 2010 A1 2016 Edition 3 1 COR 1 2019 1 IEC 61010 2 101 2018 2 101 IVD IEC 61326 1 2012 IEC 61326 2 6 2012 EMC 2 6 IVD 53 1...

Page 54: ...54 1 x1 Auto Tube x1 x1 x1 x 6 x1...

Page 55: ...2 Maelstrom 9610 95 kg 870 L x 575 W x 700 H mm 100 120Vac 50 60Hz 8A Class I 250V 10A 96 50 1600 3000 rpm 4 3900 7 UV HEPA UV 4W HEPA 55 100 120V...

Page 56: ...56 2 8 1 2 3 6 8 USB USB type A USB USB type B USB type A USB type B...

Page 57: ...57 2 USB type A USB type A PLC PLC...

Page 58: ...LED 58 2...

Page 59: ...95 3 3 1 3 3 2 3 120 4 100 120V 5 59 3...

Page 60: ...60 4 Enter 12 CK1975...

Page 61: ...61 4...

Page 62: ...62 4 24...

Page 63: ...63 4...

Page 64: ...64 Tip 4...

Page 65: ...65 8 8 8 4...

Page 66: ...66 Maelstrom 9610 25 4 1 2...

Page 67: ...67 18 1 D 18 USB 1 D 4 4 3...

Page 68: ...68 8 YES NO 4 5 6...

Page 69: ...69 RPM 1 6 4...

Page 70: ...70 4 PCR...

Page 71: ...71 Maelstrom 9610 15 4...

Page 72: ...72 UV HEPA UV UV UV UV 30 UV 30 1 2 HEPA UV HEPA HEPA UV 4...

Page 73: ...73 25 35 4...

Page 74: ...74 10 l C 25 C 100 C LB WB1 1 WB2 2 MB EB 8 4...

Page 75: ...75 35 999 9 500 3000rpm 999 9 999 9 4...

Page 76: ...76 1 2 4...

Page 77: ...77 1 2 4...

Page 78: ...78 4 Maelstrom 9610...

Page 79: ...79 Maelstrom 9610 2 1 4...

Page 80: ...80 4 1 2...

Page 81: ...81 Maelstrom 9610 USB 4 USB...

Page 82: ...82 886 3 3167568 service tanbead com 5 6 886 3 3167568...

Page 83: ...7 LED 8 83 Patent List USA US09616398B2 EU EP2937136 Canada CA2862946 Japan JP6151735B2 Korea KR101696517B1 China CN104971638B Taiwan TWI526245B WIPO WO2016127292...

Page 84: ...84 9 004831 330 81 12 2 886 3 3607555 333 188 6 886 3 3167568...

Page 85: ...85 10 M9610 UL 10A 250VAC x2 5 2 x 20mm 1 2 3 1 2 3 4 5 Fuse Fuse...

Page 86: ...86 Maelstrom Master TAN Bead Maelstrom 9610 Master Maelstrom Master Maelstrom Master TANBead Maelstrom 9610 Master USB A B Type B A B 10...

Page 87: ...87 10...

Page 88: ...88 10...

Page 89: ...100 120V Maelstrom 9610 039 M1602 X01 Rev V1 0_2021 06 11...

Page 90: ...90 1 88 2 93 3 96 4 98 5 120 6 120 7 121 8 121 9 122 10 123...

Page 91: ...91 EEA WEEE WEEE...

Page 92: ...AC 92...

Page 93: ...93...

Page 94: ...Maelstrom 9610 DNA RNA 94 1...

Page 95: ...1 120 10 35 40 80 5 60 20 90 2000 95...

Page 96: ...80 96 1...

Page 97: ...IEC 61010 1 2010 A1 2016 Edition 3 1 COR 1 2019 1 IEC 61010 2 101 2018 2 101 IVD IEC 61326 1 2012 IEC 61326 2 6 2012 EMC 2 6 IVD 97 1...

Page 98: ...98 1 x1 Auto Tube x1 x1 x1 x 6 x1...

Page 99: ...2 Maelstrom 9610 95 kg 870 L x 575 W x 700 H mm 100 120Vac 50 60Hz 8A Class I 250V 10A 96 50 1600 3000 rpm 4 3900 7 UV HEPA UV 4W HEPA 99 100 120V...

Page 100: ...100 2 8 1 2 3 6 8 USB USB type A USB USB type B USB type A USB type B...

Page 101: ...101 2 USB type A USB type A PLC PLC...

Page 102: ...LED 102 2...

Page 103: ...95 3 3 1 3 3 2 3 120 4 100 120V 5 103 3...

Page 104: ...104 4 Enter 12 CK1975...

Page 105: ...105 4...

Page 106: ...106 4 24...

Page 107: ...107 4...

Page 108: ...108 Tip 4...

Page 109: ...109 8 8 8 4...

Page 110: ...110 Maelstrom 9610 25 4 1 2...

Page 111: ...111 18 1 D 18 USB 1 D 4 4 3...

Page 112: ...112 8 YES NO 4 5 6...

Page 113: ...113 RPM 1 6 4...

Page 114: ...114 4 PCR...

Page 115: ...115 Maelstrom 9610 15 4...

Page 116: ...116 UV HEPA UV UV UV UV 30 UV 30 1 2 HEPA UV HEPA HEPA UV 4...

Page 117: ...117 25 35 4...

Page 118: ...118 10 l C 25 C 100 C LB WB1 1 WB2 2 MB EB 8 4...

Page 119: ...119 35 999 9 500 3000rpm 999 9 999 9 4...

Page 120: ...120 1 2 4...

Page 121: ...121 1 2 4...

Page 122: ...122 4 Maelstrom 9610...

Page 123: ...123 Maelstrom 9610 2 1 4...

Page 124: ...124 4 1 2...

Page 125: ...125 Maelstrom 9610 USB 4 USB...

Page 126: ...126 886 3 3167568 service tanbead com 5 6 886 3 3167568...

Page 127: ...7 LED 8 127 Patent List USA US09616398B2 EU EP2937136 Canada CA2862946 Japan JP6151735B2 Korea KR101696517B1 China CN104971638B Taiwan TWI526245B WIPO WO2016127292...

Page 128: ...128 9 330 81 12 2 886 3 3607555...

Page 129: ...129 10 M9610 UL 10A 250VAC x2 5 2 x 20mm 1 2 3 1 2 3 4 5 Fuse Fuse...

Page 130: ...130 Maelstrom Master TAN Bead Maelstrom 9610 Master Maelstrom Master Maelstrom Master TANBead Maelstrom 9610 Master USB A B Type B A B 10...

Page 131: ...131 10...

Page 132: ...132 10...

Page 133: ...MEMO...

Page 134: ...6F No 188 Wenhe Rd Guishan Dist Taoyuan City 333 Taiwan R O C Tel 886 3 316 7568 Fax 886 3 317 3369 www tanbead com Taiwan Advanced Nanotech Inc...

Reviews: