23
7. Product Performance
Consistency
The consistency of extracted lysates on the same machine and different machines (within-run
precision) is the variability of the results from well to well. The coefficient of variation of nucleic
acid extraction concentration is less than 5%. It is in the range of acceptable statistical analysis.
▓
Nucleic Acid Yield
N=48
M48LH00005
M48LH00009
M48LH00010
Average
Yield
N=48
Standard
Deviation
(SD)
Coefficient
of Variation
(CV%)
Average Yield
(ng/
l)
19.6
22.2
20.7
20.83
0.347
1.66
Standard
Deviation (SD)
0.330
0.351
0.361
Coefficient of
Variation
(CV%)
1.69
1.58
1.75
▓
Ratio (A260/A280)
N=48
M48LH00005
M48LH00009
M48LH00010
Average
Yield
N=48
Standard
Deviation
(SD)
Coefficient
of
Variation
(CV%)
Average Yield (ng/
l)
1.91
1.87
1.89
1.89
0.013
0.68
Standard Deviation
(SD)
0.0014
0.012
0.0014
Coefficient
of
Variation (CV%)
0.72
0.64
0.72