So as to prevent such a danger in advance, for testing a sample of unknown flash point, use the
Special mode in Section
3.5
. Be sure to arrange at least one proper person in your test room, lab,
etc., when testing, to cope with possible accidents.
7. Post-sales Servicing
For ordering consumable items or parts or to request servicing, please contact Tanaka authorized
distributor.
Please note that spare parts that are necessary to maintain/regain the function of this Tester is kept
in inventory for a minimum of 5 years after the delivery of this Tester.
8. Precaution in Case of Loss of Instruction Manual
The Instruction Manual is a part of this instrument. In case it is lost, please request a new copy to
Tanaka authorized distributor.
Also, it is highly recommended that the address and telephone number of Tanaka and/or its
authorized distributor be kept separately from this Manual.
9. Caution when Transferring This Tester
If this Tester is transferred or sold to a third party, please inform Tanaka or its authorized
distributor. Wrong source voltage and/or AC frequency could cause damages or malfunctioning.
In case of transferring this Tester, make sure to attach this Instruction Manual.
10. Specifications
(1)
Conforming standard: ISO 2719, ASTM D93, IP 34
(2)
Measuring Range: Ambient to 370
o
C (700
o
F)
(3)
Test Modes: ASTM D93 (ISO 2719, IP 34) A + B + C, Special (Fast Search),
Simulated TAG, User’s Custom
(4)
Display: Fluorescent display tube (VFD module)
(5)
Temperature Sensor: PT-100 in stainless steel sheath
(6)
Flash Detector: CRC Thermocouple
(7)
Heater: Band type heater 400W@100V or 400W@220V
(8)
Ignition Source: Gas ignition with automatic lighting or electric ignition, Interchangeable
(9)
Cooling Device: Forced air cooling by sirocco fan
(10)
Barometric Correction: By entering the barometric pressure reading through the
control panel, or by optional barometric pressure sensor
(11)
Safety Mechanism: Automatically shuts off and the problem is reported by
buzzer and display in case:
(a)
EFP+20
o
C (+40
o
F) or 370
o
C (700
o
F) is reached,
(b)
temperature sensor is found defective,
(c)
flash detector is found defective,
(d)
thermofuse is blown,
(e)
electric ignitor is blown,
(f)
stirring motor is stopped,
(g)
built-in battery has expired,
(h)
test cover is not set in place, or
(i)
control computer runs away (buzzer only)