8.1 PID settings
Following shows influence on stability with different PID
settings
Temperature
:
75°C
Proportional band (Pb) :
25
Integrator
:
16
Differentiator
:
0
Min/max
:
± 0.015
Temperature
:
75°C
Proportional band (Pb) :
50
Integrator
:
25
Differentiator
:
0
Min/max
:
± 0.018°C
Temperature
:
75°C
Proportional band (Pb) :
100
Integrator
:
25
Differentiator
:
0
Min/max
:
± 0.045°C
SP 75°C
0.029
± 0.015
SP 75°C
0.036
± 0.018
SP 75°C
0.089
± 0.045