INSTRUCTION FOR USE

1

Twist and pull off the cap of the 
lancing device.

2

Insert a lancet into the lancet holder 
and push down firmly until it is fully 
secured.

3

Twist the protective disk off the 
lancet.

4

Replace and twist the cap as shown in 
the picture.

1

2

1

2

www.taidoc.com

B1-7F., No.127, Wugong 2nd Rd., Wugu Dist., 
New Taipei City 24888, Taiwan
Tel : +886-2-6625-8188      Fax : +886-2-6625-0288

TaiDoc Technology Corp.

Copyright TaiDoc Technology Corporation. All rights reserved.

V2

9

Remove the lancet by pulling the cap 
off first. Safely dispose of the used 
lancet by placing the protective disk 
on a hard surface, and pushing the 
exposed tip into the disk. 

10

Slide the ejector forward to remove 
the used lancet. Dispose of the lancet 
according to your local regulations.

5

Select the depth of penetration by 
turning the adjustable tip in either 
direction so that the arrow on the cap 
points to the desired depth.

6

Pull the cocking control back until it 
clicks. You will see a color change 
inside the release button when it is 
ready. 

7

Lancing the finger
Press the lancing device’s tip firmly 
against the lower side of your 
fingertip. Press the release button to 
prick your finger, then a click 
indicates that the puncture is 
complete.

8

Remove the device from the puncture 
site. After penetration, remove the first 
drop of blood with a clean cotton swab. 
Gently squeeze the punctured area to 
obtain another drop of blood. Be careful 
NOT to smear the blood sample. Apply 
the blood to the test strip.

Hold your fingers on the
body NOT the cap. 