background image

2

Light settings

To get the best results, we recommend using this application with the 
TAGARNO FHD Trend. This allows you to attach the right accessories 
and rightfully adjust the settings.

Equipment needed:

Ring light white
Basic or Advanced control box (external lighting) 
White balance calibration card
White balance calibration card fixture
Sample drawer
Sample drawer positioner
Light eliminator

This enables the app to run objective and reproducable analyses.

The microscope light settings are automatically set to auto exposure.   

Step 1

Mount a Ring light white from TAGARNO on the microscope as 

described in the TAGARNO Ring light white user guide. 

Step 2

Adjust the light settings to the right intensity using the advanced ring 

light control box.

Step 3

Place the Light eliminator over the camera head as displayed below 

and close the Light elimator by pulling the backwall all the way up. Do 

not forget to tighten the screws to prevent the backwall from falling 

back down. Using the Light eliminator will eliminate any external light 

and prevent it from interfering with the analysis result.  

Step 4

Run a white balance calibration (see Toolbar section, pt 4).  

Step 5

Pour sample into the Sample drawer, insert it in the Sample drawer 

positioner and pull the Light eliminator all the way down using the 

handles as shown below. 

1.  ASSEMBLING