5
6.
8.
1.
2.
7.
3.
4.
5.
6. CONNECTING
Connect USB 2.0 (see options below) (1+2), control box (3),
monitor (4), ethernet (5), USB 3.0 (6), ESD (Only black models) (7)
and power supply (8) in that order.
USB 2.0 (1+2): Keyboard, mouse, USB key, USB hard drive, scanner
Connect keyboard and mouse in the USB 2.0 inputs (only for use
with some applications). Keyboard and mouse need to be HID
compliant.
When more than 2 USB plugs are needed, use a USB hub. If you
have received a USB hub with your purchase, remove the foil,
connect it to one of the USB ports and attach it to the back of the
microscope.
Connector Description
Type
Connect to
Specification
1+2
USB 2.0 input
USB 2.0 Type A Socket Keyboard
Mouse
USB key
USB hard drive
Barcode scanner
USB 2.0 Host. Output: 5V 500mA
3
Control box
D-SUB 9-Pin Female
Control box
Output: 3.3V
6A Max
Use only control boxes supplied by TAGARNO
4
Output
HDMI Standard Type A Monitor
HDMI Out 1080p60
5
Ethernet
RJ-45
Ethernet wall socket
Ethernet LAN, 100BASE-TX/1000BASE-T
6
USB 3.0 output
USB 3.0 Type B Socket Computer
USB 3.0 Device. Self-powered
7
ESD
Snap Socket, 10mm
ESD Grounding point
ESD Grounding point. Only black models
8
Power supply
(DC)
DC Jack
Power socket
Input: 12V 1.7A
Use only the power supply provided by TAGARNO:
Brand/Model: Mean Well/GSM40A12
Protection Class I
Input: 100-240V
1.0-0.5A, 50/60Hz
Overvoltage category ll
Output: 12V
3.34A Max
HDMI - HDMI cable
Live image
The microscope can be connected directly to a monitor.
Live image and image capturing through
USB 3.0 /
3rd part software.
See system recommended (third-party) under system requirements.
You are still able to connect monitor(s) as described above.
NB:
Do not connect the HDMI cable to a PC.
Live image
/ image capture
USB 3.0
Ethernet cable
D-SUB
Ethernet cable (or WiFi)