2 Installation
2.3 Sample Conditioning System
The sample-conditioning system ensures a clean, regulated sample for the analyzer. Sample-gas
composition and pressure, along with other factors, may require modifying the conditioning system
for a specific application. Mount the sample-conditioning system below the analyzer if possible. The
typical system may consist of a sample probe, drip pot, one or two regulators, and a membrane filter.
Some samples require an additional pressure regulator at the sample tap. One to three pressure drops
may be necessary to ensure that the analyzer receives a sample that is free of liquids. A typical system
is illustrated below.
NOTE: Configuration may be different pending the analyzer ordered.
Plate Dimensions: H=11" x W=7"
Sample In
(1/4" – not to exceed 400 PSIG)
PIPE LINE
3/8"
Sample
Out
(1/4")
+ / -
+ / -
+ / -
+ / -
+
FRONT VIEW
SAMPLE SYSTEM
to back
of panel
Important:
To keep
liquids from reaching
the analyzer, sample
plate
MUST
be mounted
BELOW
the analyzer.
Tip:
Position the
solar panel as a sun
shade to prevent
glare on the PLC.
Ball Valve
(recommended)
Tubing
(client supplied)
Low Pressure In
Probe
(recommended)
Liquid Stop
Low
Pressure
Out
Membrane Filter
Tubing
(client supplied)
Needle Valve
(leave slightly open)