Page 52
AR-G2/AR 2000ex/AR 1500ex Getting Started Guide
Error & LCD Display Messages
The LCD display on the front of the rheometer electronics box displays useful information and error
messages.
Power On Messages
Immediately after power up of the rheometer, the display will show
'Initialising...'
After a few seconds
this will change to
'AR-G2 8.xx xx/xx/xx' (x is version dependent)
. If it does not, there is a problem
with the rheometer.
After a few more seconds, the display will either show
'System test Ok'
or
'System test failed Y,'
where
Y is an error code as show below:
Initialising ...
During start up of the rheometer, the following items are shown as initializing:
•
Electronics
•
Power board
•
Instrument
Bearing overspeed
Shown when the rheometer bearing rotation exceeds the specified maximum speed.
Bearing pressure too low
This is displayed if the air supply has been inadvertently switched off while the rheometer was on or if the
supply pressure has dropped below the minimum operating pressure.
1
ROM checksum error
Either bad firmware or hardware
fault
Try reloading firmware,
otherwise call for service.
2
RAM error
Hardware fault
Call for service.
4
Dual port RAM error
Hardware fault
Call for service.
40
Battery failure
Low battery
Sometimes seen the first time the
system is restarted after a firm-
ware upgrade. Cycles power off
then on
80
Backup RAM error
Hardware fault
Call for service.
400
Parameter block
checksum error
Either corrupted internal system
parameters or hardware fault
Call for service.
Summary of Contents for AR-G2
Page 1: ...Revision R Issued January 2011 Getting Started Guide AR G2 AR 2000ex AR 1500ex Rheometers...
Page 12: ...Page 12 AR G2 AR 2000ex AR 1500ex Getting Started Guide This page is intentionally blank...
Page 16: ...Page 16 AR G2 AR 2000ex AR 1500ex Getting Started Guide This page is intentionally blank...
Page 54: ...Page 54 AR G2 AR 2000ex AR 1500ex Getting Started Guide This page is intentionally blank...