Choosing a Location
It is important to choose a location for the instrument using the following guidelines. The RSA-G2 should
be:
In
• A temperature- and humidity-controlled area. Temperatures should be in range of 20 to 30°C.
• A clean, vibration-free environment, preferably on the ground floor in the building. It should be
located away from pumps, motors, or other devices which produce vibrations.
• An area with ample working and ventilation space.
• An area that allows at least six inches of clearance on all sides of the equipment, and that allows a min
-
imum of eight inches from the open and closed positions of all heated environment control systems.
• An area that does not restrict airflow to cooling fans on the rear of the power control box, or the option
controllers.
On
• A stable work surface.
A marble table is recommended
. Isolation mounts on a standard lab bench are
not recommended.
Near
• A power outlet. An RSA-G2 with an oven requires 180 to 264 VAC/50 or 60 Hz with a 20 amp outlet.
• Your TA Instruments TRIOS Instrument Control Software.
• Compressed lab air and purge gas supplies with suitable regulators and filters.
Away from
• Dusty environments.
• Exposure to direct sunlight.
• Direct air drafts (fans, room air ducts).
• Poorly ventilated areas.
• Noisy or mechanical vibrations.
• High traffic areas, where constant movements from passing personnel could create air currents or
mechanical disturbances.
Summary of Contents for RSA-G2
Page 1: ......
Page 9: ...Forced Convection Oven with tension clamp Technology 8...
Page 13: ...IMMERSION TESTING I CLAMPING SYSTEMS...
Page 15: ...DETA I DIELECTRIC THERMAL ANALYSIS ACCESSORY DETA...
Page 29: ...tainstruments com 2019 TA Instruments All rights reserved L50027 001...
Page 30: ...Verso Filler Page z z...
Page 31: ...n getting Started guide Revision D Issued July 2022...
Page 99: ...Issued November 2013 Supplement to Getting Started Guide TA Instruments Manual Supplement...
Page 100: ...Page 2 TA Instruments Manual Supplement This page is intentionally blank...
Page 113: ...Page 1 Page 1 RSA G2 Solids Analyzer Revision G Issued May 2022 Site Preparation Guide...