Connect a power cable to the connector below the power switch, but do not connect this to the main
power supply at this point.
6
Install any optional upper power supply units, such as the forced convection oven or environmental
control power supplies.
Forced Convection Oven Controller Power Supply (used in conjunction with an
FCO).
CAUTION: Make sure there is no power going to the power supply unit while making the connec-
tions described here. Turn the power switch to the off position.
MISE EN GARDE: Assurez-vous que le bloc d'alimentation électrique n'est pas alimenté en courant
pendant le réglage des connexions décrites ici. Placez l'interrupteur d'alimentation sur la position
d'arrêt.
To connect the FCO power supply, follow these steps:
1
Place the FCO controller power supply on top of the test station (lower) power supply or on top of a
previously-installed environmental control power supply within the limit of the cable length. Make sure
you have access to the rear of the power supply and the ARES-G2. See the figure below.
2
Obtain the loose end of one of the accessory cables that have already been connected to the lower power
supply (see
on the previous page), and insert it into the accessory port on the FCO power supply.
Figure 29
FCO Controller power supply.
3
Oven
: Connect the power cable from the oven to the connector to the left of the fan.
4
Obtain the loose end of one of the power cables that are already connected to a switched accessory power
outlet on the lower power supply (
in the previous section). Connect this to the power inlet next
to the power switch on the FCO power supply.
5
Using the grip handles, position the power supplies as desired. (They may be placed on the floor or on
the table top.)
6
Proceed to the next sections to connect the LN2 accessory and/or another power supply box, if applicable.
DO NOT switch on the power supply at this point.
Power
switch
Accessory
power cord to
switched mains
Power
cable
to oven
Fan
Power
cable to
LN2 Cooler
Accessory
to test station
power supply
communication