SECTION 10 CALIBRATION                                   

Press 

UNIT

  key  and 

M+

  key  together  when  normal  weighing  mode,  display  shows 

F0 H-L

,  press 

TARE

  key  until  display  shows 

PrOG

,  press 

ZERO

  key,  display 

shows 

PIN

,  You  can  press 

G/N

UNIT

ZERO

  key  to  enter  setting  mode,  press 

TARE

  key  to  select  "

p2

 

CAL

",  press 

ZERO

  key  to  sure.  press 

TARE

  key  to 

select  .While  it  is  showing  "

CAL

      "  press  the 

ZERO

  key  to  enter.  The  display  will 

show    "

linear

". 

1

.

 Normal calibrate

 

Press 

TARE

 key to select    "

nonlin

".then press the 

ZERO

 key to enter,the display 

shows:  "

unload

  Remove  any  weight  from  the  platform.    After  stable  indicator  on, 

press the 

ZERO

 key. 

 
Then the display will show the last calibration weight used.    If this is correct ,you can 
continue  by  pressing  the 

ZERO

  key.      If  it  is  not  correct  use  the 

M+,

 

G/N

TARE 

keys to change the calibration weight value.    When it is correct press the 

ZERO

 key. 

 
Then  display  will  show 

LoAd

.    Place  the  calibration  weight  on  the  scale.  After 

stable, press the 

ZERO

 key. Then the calibration has completed. 

 

2. Linear calibrate 

Press the 

ZERO

 key to enter,the display shows: "

ld 0

, Remove any weight from the 

platform.    After stable indicator on, press the 

ZERO

 key. 

 
Then  display  will  show 

Ld 1

.    Place  the  1/3  of  capacity  calibration  weight  on  the 

scale. After stable, press the 

ZERO

 key. 

 
Then  display  will  show 

Ld 2

.    Place  the  2/3  of  capacity  calibration  weight  on  the 

scale. After stable, press the 

ZERO

 key. 

 
Then display will show 

Ld 3

.    Place the full capacity calibration weight on the scale. 

After stable, press the 

ZERO

 key. Then the calibration has completed. 

 
If the calibration is acceptable the display will return to normal.    If an error message 
is  shown  try  calibration  again  as  a  disturbance  may  have  prevented  a  successful 
calibration.     
 
If the problem persist then contact your dealer. 
 
After calibration the scale should be checked to verify the calibration and linearity is 
correct.    If  necessary  repeat  calibration,  especially  be  certain  the  scale  is  stable 
before accepting any weight. 

                                                                                                                                                                    -16- 

Summary of Contents for ROW Series

Page 1: ...ROW series Precision Balance User s guide UGROW E1 01...

Page 2: ...ION 4 KEY DESCRIPTIONS 5 SECTION 5 DISPLAYS 6 SECTION 6 OPERATION 7 6 1 Zeroing the display 7 6 2 Taring 7 6 3 Weighing a sample 7 6 4 Check weighing 8 6 5 Accumulated total 9 SECTION 7 PARAMETERS 10...

Page 3: ...nd check weighing functions All the keypads are sealed color coded membrane switches and the displays are large easy to read liquid crystal type displays LCD The LCD s are supplied with a backlight Al...

Page 4: ...0 C 40 C 32 F 104 F Power supply external 115 230 Vac 120 240 Vac 50 60Hz 10 watts Calibration Automatic External Display 6 digits 22mm LCD display attached backlight indicator Housing ABS Plastic Ze...

Page 5: ...humidity that might cause condensation Avoid direct contact with water Do not spray or immerse the scales in water Avoid air movement such as from fans or opening doors Do not place near open windows...

Page 6: ...ther functions G N Press the key the scale can to select gross weight or net weight after you tare a weight Secondary function In the setting mode this key used to move active digits right M To print...

Page 7: ...ser s manual SECTION 5 DISPLAYS The LCD display will show a value and a unit to the right of the digits In addition there are labels for TARE GROSS weight Zero and for Low battery 5 Battery low stable...

Page 8: ...lue and that value is subtracted from the display leaving zero on the display The TARE indicator will be on As product is added only the weight of the product will be shown The scale could be tared a...

Page 9: ...RO key to enter press TARE key until display show BEEP press ZERO key to enter press TARE key to select BP 2 check mode 2 BP3 check mode 3 BP1 no beep press ZERO key to sure press UNIT key to escape 6...

Page 10: ...e mode Press the UNIT key and M key together in the weighing mode it will display F0 H L press TARE key until display show f5 prt press ZERO key to enter press TARE key to select p auto press ZERO key...

Page 11: ...detail in SECTION12 Turn on power scale will show err4 then show current total weight on platform press TARE key display show 0 00 then remove goods in hopper display will show it s weight in mode pre...

Page 12: ...e TARE to skip to the next parameter Set Hi for setting the high limit After pressing the ZERO key to set a limit use the G N keys to change the flashing digit then use the TARE key to increment the f...

Page 13: ...weighing data to PC Command T do tare operate Command Z do zero operate P auto auto M mode Sei re also send continuously Then set baud rate After set working mode display will show b xxx this is curre...

Page 14: ...11 11 11 GS 0 222kg TOTAL 0 888kg DATE 04 06 06 TIME 11 11 11 NT 0 222kg TW 0 666kg GW 0 888kg TOTAL 0 888kg 4 NO 4 GS 0 888kg NO 4 NT 0 666kg TW 0 222kg GW 0 888kg NO 4 GS 0 222kg TOTAL 1 000kg No 4...

Page 15: ...attery simply plug into the mains power The scale does not need to be turned on The battery should be charged for 12 hours for full capacity In the LCD display has an battery indicater to indicate the...

Page 16: ...he Total weight stored in the accumulation memory the output format is 9 3 continuously output protocol con1 weighing mode 14 Date AND TIME The scale will be set date and time S N The number increment...

Page 17: ...et mode tare value not 0 header2 header2 01H If gross weight header2 header2 02H If overload or gross weight header2 header2 04H If unstable header2 header2 08H If weighing unit kg header2 header2 10H...

Page 18: ...show LoAd Place the calibration weight on the scale After stable press the ZERO key Then the calibration has completed 2 Linear calibrate Press the ZERO key to enter the display shows ld 0 Remove any...

Page 19: ...es Format hh mm ss Err 4 Zero Setting Error The scale was outside the normal zero setting range either when it was turned on or when the ZERO key was pressed Remove weight from the scale and try again...

Page 20: ...ADC speed press Tare key to enter 7 5 7 5 times per second 15 15 times per second 30 30 times per second 60 60 times per second Note 15 times per second or 30 times per second are recommendatory DECI...

Page 21: ...ROW serial precision balance user s manual MODE 4 As the mode 2 but M out format different NW 0 222KG nw for net weight GW 0 888KG gw for gross weight 18...

Page 22: ...their distributors offer a full range of technical services such as on site and workshop repair preventative maintenance and calibration facilities TAIWAN SCALE operates an approved quality managemen...

Reviews: