SECTION 10 CALIBRATION                                     

 

10.1 Linear calibrate 

To start calibration turn the scale off and then turn it back on.    Press the 

limit

 and 

%

 

keys together during the initial counting from 9 to 0 on the display. 
 
The display will show "unLoAd".    Remove all weight from the pan and then press the 

ZERO/ENTER

 key to set the initial zero point. 

 
The display will then show the first calibration weight request.    Put this weight on the 
platform and then press the 

ZERO/ENTER

 key    when stable.    The scale should be 

stable before pressing the 

ZERO/ENTER

 key to accept a weight.  The stability 

indicator will turn on to show the value is stable. 
 
After all calibrate weight completed, press 

ZERO/ENTER

 to complete the calibration.   

The scale will begin counting back to zero after the last weight has been selected. 
 
Calibration weights   

QHW SERIES 

Model # 

QHW 3 

QHW 6 

QHW 15 

QHW 30 

Weight 

1  zero zero zero zero 

Weight 2 

1000g 

2kg 

5kg 

10kg 

Weight 3 

3000g 

6kg 

15kg 

30kg 

 

10.2 Normal calibrate 

To start calibration turn the scale off and then turn it back on.    Press the 

Tare

 and 

%

 

keys together during the initial counting from 9 to 0 on the display. 
 
The display will show "unLoAd".    Remove all weight from the pan and then press the 

ZERO/ENTER

 key to set the initial zero point. 

 
Then use 

%

Limi

t and 

Tare

 key to key in calibrate weight, press 

Zero

 key to sure, 

display shows “load”, put this weight on platter, after stable indicator on, press 

Zero

 

key to sure. 
 
After selfchecking again, QHW is ready for you. 
 
 
 
 
 
 
 
 
 

 
 
 
 
 
 
 

Summary of Contents for QHW 15

Page 1: ...UGQHW E0204 QHW series Digital Weighing Scales User s guide...

Page 2: ......

Page 3: ...5 DISPLAYS 5 SECTION 6 OPERATION 6 6 1 Zeroing the display 6 6 2 Taring 6 6 3 Weighing a sample 6 6 4 Percent weighing 6 6 5 Parts counting 7 6 6 Check weighing 7 6 7 Accumulated total 8 6 8 Manual a...

Page 4: ......

Page 5: ...They all have stainless steel weighing platforms on an ABS base assembly All the keypads are sealed color coded membrane switches and the displays are large easy to read liquid crystal type displays L...

Page 6: ...ns Interface RS 232 Output Optional Stabilisation Time 2 Seconds typical Operating Temperature 0 C 40 C 32 F 104 F Power supply external 115 230 Vac 50 60Hz 10 watts Calibration Automatic External Dis...

Page 7: ...ater Avoid air movement such as from fans or opening doors Do not place near open windows Keep the scales clean Do not stack material on the scales when they are not in use 3 2 INSTALLATION of QHW SER...

Page 8: ...the right when setting values for parameters or other functions Enters the percent weighing function Allows the weight unit weight and count to be seen when parts counting Secondary function is to mo...

Page 9: ...and a unit to the right of the digits In addition there are labels for TARE GROSS weight ZERO and for Low battery 888888 OK HI LO GROSS TARE ZERO g pcs oz tkglb Checkweighing Symbols Stability Symbol...

Page 10: ...time if another type of product was to be added to the first one Again only the weight that is added after taring will be displayed When the container is removed a negative value will be shown If the...

Page 11: ...to display unit weight g pcs Total weight kg or the count pcs Press the FUNC key to return to normal weighing 6 6 CHECK WEIGHING Check weighing is a procedure to cause an alarm to sound when the weigh...

Page 12: ...he zero values 6 7 ACCUMULATED TOTAL The scale can be set to accumulate automatically when a weight is added to he scale or manually by pressing the PRINT key See the PARAMETERS Section for details of...

Page 13: ...s to zero Adding a second weight will repeat the process While the weight is on the scale it is permissible to press the PRINT key to store the value immediately In this case the scale will not store...

Page 14: ...y switch off to protect the battery To charge the battery simply plug into the mains power The scale does not need to be turned on The battery should be charged for 12 hours for full capacity Just und...

Page 15: ...ormal weighing operations parts counting or recalling of totals from memory will all be different Examples follow Normal Output When percent weighing the output is the weight shown in percent only GS...

Page 16: ...g off when the scale is not used The backlight can be set to be EL On EL Au Automatic or EL Off The maximum battery life is achieved with the backlight turned off Press the TARE key to change the valu...

Page 17: ...l calibrate weight completed press ZERO ENTER to complete the calibration The scale will begin counting back to zero after the last weight has been selected Calibration weights QHW SERIES Model QHW 3...

Page 18: ...ON POSSIBLE CAUSES Err 4 Initial Zero is greater than allowed 4 of maximum capacity when power is turned on or when the ZERO ENTER key is pressed Weight on the pan when turning the scale on Excessive...

Reviews: