IP30 Label Printing Scale Operation Manual
- 45 -
Disconnect battery and charger, connect again
and turn on the power
Disconnect AC
adapter will turn
off power
Battery low
Verify the batteries are rechargeable types and
charging voltage getting proper. Plug AC
adapter to charge battery.
Replace battery
LCD off but
power LED on
Scale in sleep mode
System hang
Press ON/OFF key to back to working mode
Hold ON/OFF key for 6 seconds to restart the
scale.
Disconnect battery and charger, connect again
and turn on the power
Can
’
t turn on
power after
update software
by USB
System hang
Update software incorrect,
Used old core software.
Disconnect battery and charger, connect again
and turn on the power.
Use another USB flash driver to update
software.
Update core software.
Not Printing
Printer Format not select,
Select the print format as default or for the
specific product .
7.1. Error messages
--ol--
Over load
Remove the load.
Re calibrate
--Ul--
Underload
Minus weight, check the platform and restart or
calibrate.
Err 4
Zero setting error
Zero setting range exceeded due to switching
on.(4%max)
Make sure platform empty.
Err 19
Initial zero setting error
Calibrate the scale
In the case of errors which you cannot eliminate yourself and which are not described in
these operating instructions, please contact T-Scale technical support team.
Summary of Contents for IP30
Page 1: ...IP30 Label Printing Scale Operation Manual A1 Manual_IP30_V1...
Page 9: ...IP30 Label Printing Scale Operation Manual 6 2 7 Customer Display...
Page 13: ...IP30 Label Printing Scale Operation Manual 10 3 5 Cassette Label Printer 2...
Page 14: ...IP30 Label Printing Scale Operation Manual 11 Replacement of Label Roll...
Page 16: ...IP30 Label Printing Scale Operation Manual 13...
Page 51: ......