SZBJ BM801A Instruction Manual Download Page 2

4.3 DC Voltage Measurement 

 

 

WARNING!

 Maximum Input Voltage is 1000V AC/DC. 

Do  not  attempt  to  take  any  voltage  measurement  that  may 
exceed  to  avoid  electrical  shock  hazard  or  damage  to  this 
meter.   
(1)

 

Set the rotary switch at DC1000V position. Connect  the 
red  test  lead  to  “V/

”  jack  and  the  black  test  to  the 

“COM” jack. 

(2)

 

Connect  test  leads  across  the  source  or  loading  being 
measured and read voltage value on the LCD display. 

 
4.4 AC Current Measurement 

 

 

WARNING!

  The  input  current  should  not  exceed 

indicated  values,  it  is  to  avoid  electrical  shock  hazard  or 
damage to this meter.   
(1)

 

Set the rotary switch at AC1000A position. 

(2)

 

Press the trigger to open transformer jaw and clamp  one 
conductor only, making sure that the jaw is firmly closed 
around  the  conductor.  Read  current  value  on  LCD 
display. 

(3)

 

If current under test is smaller, the lower range has to be 
selected.   

Note: 

a)  Disconnect  the  test  lead  with  the  meter  for  safety 

before this measurement.   

          b)  If  two  or  more  conductors  are  clamped,  the  meter 

will stop operating.     

          c)  The  accurate  reading  will  be  obtained  by  the 

conductor across center of the transformer jaws. 

   
4.5 Resistance Measurement 

 

 

WARNING! 

When  checking  in-circuit  resistance,  be 

sure  the  circuit  under  test  has  all  power  removed  and  all 
capacitors have been discharged fully. 
(1)

 

Set the rotary switch at 2k

 position. 

(2)

 

Connect the red test lead to “V/

” jack and the black test 

to the “COM” jack. 

(3)

 

Connect test leads across the resistor to be measured and 
read LCD display. 

(4)

 

When at open circuit or over-range condition, “1” will be 
displayed. 

 
4.6 Diode Measurement 

(1)

 

Set the rotary switch at 

position. When the input is 

not connected, i.e. at open circuit, the figure “1” will be 
displayed. 

(2)

 

Connect the red test lead to “V/

” jack and the black test 

to  the  “COM”  jack.(The  polarity  of  red  lead  is  positive 
“+”) 

(3)

 

If diode test mode is selected, connect the red and black 
leads to the anode and cathode of the diode, the forward 
voltage drop of this diode will be displayed in mV. 

(4)

 

The  figure  “1”  will  be  displayed  when  the  diode  is 
reversed under test. 

 
4.7 Continuity Test 

(1)

 

Set the rotary switch at 

position. When the input is 

not connected, i.e. at open circuit, the figure “1” will be 
displayed. 

(2)

 

Connect the red test lead to “V/

” jack and the black test 

to the “COM” jack. 

(3)

 

Connect  tip  of  the  test  leads  to  the  points  where  the 
conduction  needed.  If  the  resistance  is  under  50

,  the 

buzzer will sound continuously. 

Note:

 The circuit to be tested must in power off status during 

the continuity check, because any load signals may sound the 
buzzer and mislead the result. 
 
 

5. Maintenance 

 

 

WARNING!

  To  avoid  electrical  shock  hazard,  before 

open  the  case  or  the  battery  cover,  be  sure  that  test  leads 
have  been  disconnected  from  measurement  circuits.  Close 
case and tighten screws completely before using the meter. 

5.1 If the sign “

” appears on the LCD display, it indicates 

that battery should be replaced. Remove the battery cover of 

case, replace the exhausted battery with a new one. 
5.2 Do not expose the instrument to direct sunlight, extreme 
temperature or moisture. 
5.3  Keep  the  meter  cleaning.  Do  not  use  abrasives  or 
solvents  on the  meter, wipe  periodically the case  with a dry 
cloth and detergent.     
 

Reviews: