background image

D200-09-00 

I56-1986-000

C0169-00

C0170-00

Cleaning

It  is  recommended  that  the  detector  be  removed  from 

its  mounting  base  to  facilitate  cleaning.  The  detector  is 

cleaned as follows:

NOTE:

  Before  removing  the  detector,  notify  the  proper 

authorities  that  the  smoke  detector  system  is 

undergoing  maintenance  and  will  be  temporarily 

out of service. Disable the zone or system under-

going maintenance to prevent unwanted alarms.

1.  Remove the detector cover by prying away the four 

side  tabs  with  a  small-bladed  screwdriver,  and  then 

pulling the cover from the base.

2.  Vacuum  the  screen  carefully  without  removing  it.  If 

further cleaning is required continue with Step 3, oth-

erwise skip to Step 7.

3.  Remove the screen/chamber cover assembly by pull-

ing it straight out (see Figure 4).

4.  Clean the chamber by vacuuming or blowing out dust 

and particles.

5.  Replace  the  sensing  chamber  cover,  aligning  the 

arrow  on  the  top  with  arrow  on  the  printed  circuit 

board.

6.  To  replace  the  screen,  place  it  over  the  chamber 

assembly, turning it until it snaps into place.

7.  Replace the cover using the LEDs to align the cover 

and then gently pushing it until it locks into place. 

8.  Reinstall the detector.

9.  Test the detector as described in TESTING.

10.  Reconnect disabled circuits.

11.  Notify the proper authorities that the system is back 

on line.

C0146-00

Figure 2. Rotary decade address switches:

Figure 3. Test magnet position:

TENS

ONES

9

8

7

6

5

4

3

2

1 0

9

8

7

6

5

4

3

2

1 0

D

N

T

N

I

A P

T

O

O

LED STATUS

INDICATORS

TES

T

MAGNET

PAINTED

SURFACE

PAINTED

SURFACE

TEST

MAGNET

MAGNET TEST

MARKER

Figure 4. Sensor assembly:

SENSING

CHAMBER

SCREEN

ASSEMBLY

SENSOR

COVER

COVER REMOVAL

TABS

Summary of Contents for 7251A

Page 1: ...perating Voltage Range 15 to 32 VDC Standby Current 330µA 24 VDC one communication every 5 sec with LED blink enabled Max Alarm Current LED on 6 5 mA 24 VDC Operating Humidity Range 10 to 93 Relative Humidity noncondensing Operating Temperature Range 0 C to 38 C 32 F to 100 F Height 1 7 inches 43 mm installed in B210LPA Base Diameter 6 1 inches 155 mm installed in B210LPA Base 4 1 inches 104 mm in...

Page 2: ...in the magnet test area as C0100 00 3 2 1 3 2 1 OPTIONAL RETURN LOOP RE M OTE ANNUN CIATOR shown in Figure 3 2 The sensor should alarm the panel Two LEDs on the sensor are controlled by the panel to indicate sensor status Coded signals transmitted from the panel can cause the LEDs to blink latch on or latch off Refer to the control panel technical documentation for sensor LED status operation and ...

Page 3: ...y by pull ing it straight out see Figure 4 4 Clean the chamber by vacuuming or blowing out dust and particles 5 Replace the sensing chamber cover aligning the arrow on the top with arrow on the printed circuit board 6 To replace the screen place it over the chamber assembly turning it until it snaps into place 7 Replace the cover using the LEDs to align the cover and then gently pushing it until i...

Page 4: ...ich case the user will be required to correct the interference at his own expense Laser Safety Information This smoke detector does not produce any hazardous laser radiation and is certified as a Class 1 laser product under the U S Department of Health and Human Services DHHS Radiation Performance Standard according to the Radiation Control for Health and Safety Act of 1968 Any radiation emitted i...

Reviews: