SYSMEX XW-100 Clinician'S Quick Manual Download Page 2

2

XW-100 CLINICIAN'S INFORMATION

This Quick Guide is for use with the Sysmex XW-100. Your facility must:

• 

Have a CLIA Certificate of Waiver in order to perform waived tests.

• 

Meet all applicable state and local laboratory testing laws.

• 

Follow the manufacturer's instructions; please read the complete test procedure before performing a test.

WBC

• 

Fragmented RBCs

• 

Platelet aggregation

• 

Cold agglutinin

• 

Inclusion of white blood cells

Sample Flags and Possible Causes

• 

Lyse resistant RBCs

• 

Immature white blood cells

• 

White blood cell agglutination

• 

Platelet satellite phenomenon

• 

Hyper segmented neutrophils

• 

Cryoglobulins

Flag

Possible Causes

• 

Nucleated RBCs

• 

Malaria parasites

• 

Fragile white blood cells

• 

Platelet aggregation

• 

Elevated WBC count

• 

Sample coagulation

RBC

• 

Nucleated RBCs

• 

In vitro hemolysis

• 

Sample coagulation

• 

Giant platelets

PLT

• 

Platelet aggregation

• 

Non-uniformity in platelet size

• 

Cytoplasmic fragments of 
nucleated cells

• 

Low platelet count

• 

Fragmented RBCs

• 

Cold agglutinins

• 

Sample coagulation

WBC 
Diff

• 

Lyse resistant RBCs

• 

Increased mixed population

• 

Decreased neutrophils

• 

Immature granulocytes

• 

High values for monocytes

• 

High values for Eosinophils

• 

High values for Basophils

• 

Aged sample

• 

CML

• 

Giant platelets

• 

Microcytes

• 

Cryoglobulins

• 

Lipids

• 

Microcytes

• 

Anisocytosis

• 

High WBC

• 

Giant platelets

• 

Aged sample

• 

Fibrin

• 

Proteins

• 

Lipids

Introduction

The XW-100 Automated Hematology Analyzer (XW-100), is a quantitative automated hematology analyzer 
intended for 

in vitro diagnostic use to classify and enumerate the following parameters for venous whole blood 

anti-coagulated with K2/K3 EDTA: WBC, RBC, HGB, HCT, MCV, PLT, LYM%, Other WBC%, NEUT%, LYM#, 
Other WBC#, NEUT#. It is not for use in diagnosing or monitoring patients with primary or secondary chronic 
hematologic diseases/disorders, oncology patients, critically ill patients, or children under the age of 2.

Summary of Contents for XW-100

Page 1: ...EMATOLOGY ANALYZER CLIA WAIVED CLINICIAN S QUICK GUIDE BEYOND A BETTER BOX Code No AS693925 en am PRINTED IN JAPAN Date of Last Revision 05 2018 Document Version 2 1 Software Version Ver 1 SYSMEX CORP...

Page 2: ...telets PLT Platelet aggregation Non uniformity in platelet size Cytoplasmic fragments of nucleated cells Low platelet count Fragmented RBCs Cold agglutinins Sample coagulation WBC Diff Lyse resistant...

Page 3: ...Adolescents 12 to 21 years of age 3 0 10 0 25 0 100 0 3 0 4 7 4 20 10 0 11 9 25 0 36 9 100 0 162 0 1 9 40 0 0 4 15 0 80 0 4 8 10 8 4 20 6 10 12 0 18 0 37 0 52 0 163 0 369 0 1 9 8 6 40 0 80 0 0 4 3 9 1...

Page 4: ...convenient work space Near an internet connection provided Ethernet cable is 6 6 feet long and a 3 prong power outlet provided power cord is 6 6 feet long Ensure the space is level dry dust free and...

Page 5: ...the Ethernet Cable plug into the network socket 9 During Initial Setup the system will instruct the user to scan the barcode on the Quick Reference Guide Use the barcode below to do so If the Barcode...

Page 6: ...posal cycle to prevent the spread of infection 1 Put on disposable gloves eye protection and lab coat 2 Remove the waste cap and empty or discard the waste container following your laboratory s rules...

Page 7: ...100 reagents XW CELLCLEANTM and XW QC CHECK must be stored at room temperature 59 to 77 F or 15 to 25 C Store XW CELLCLEAN in the box it was shipped in until use Keep the extra reagent XW CELLCLEAN a...

Page 8: ...un within 8 hours of having been drawn if stored at room temperature or within 36 hours if refrigerated Required sample volume 1 mL Instrument Preparation If the screen displays QC Required refer to t...

Page 9: ...children under 2 years of age must not be tested 2 Insert the white adapter when prompted 3 Enter the Operator ID 4 Confirm the Operator ID 5 Enter the Patient ID 6 Confirm the Patient ID 7 Enter the...

Page 10: ...Review Results IMPORTANT Deliver the printed test results promptly to the person who ordered the test All test results must be interpreted by a clinician The thermal paper printout is NOT permanent th...

Page 11: ...ONENTS Power Cord Waste Tubing Ethernet Cable Container Spout Kit XW pack D XW pack L Reagent Tray XW 100 Green Adapter White Adapter Thermal Paper Barcode Reader Waste Bottle XW pack D XW pack L XW C...

Page 12: ...n a second screen displays 2 Press the Move Instrument button this will cause the reagents to drain and the XW 100 to shut down 3 Follow the on screen instructions you may be prompted to press the Pow...

Page 13: ...EMC Information This instrument complies with IEC61326 2 6 2012 EMI Electromagnetic Interference For this standard the requirements of class B are fulfilled EMS Electromagnetic Susceptibility For this...

Page 14: ...non shielded equipment interface cable to this equipment will invalidate the IEC Certification of this device and may cause interference levels which exceed the limits established by the IEC for this...

Page 15: ...15 NOTES...

Page 16: ...Programs and specification subject to change without notice Sysmex America Inc 577 Aptakisic Road Lincolnshire IL 60069 U S A Tel 1 800 379 7639 www sysmex com us Sysmex Corporation 1 5 1 Wakinohama...