Manual and Mechanical Cleaning.
Instructions for the Synthes Small Battery Drive II.
1 Remove debris.
Rinse the device under running cold tap water for a
minimum of 2 minutes. Use a sponge, soft lint-free cloth
or soft-bristled brush to assist in removing gross soil.
For cannulations of the handpiece and attachments, the
cleaning brush (519.40) should be used.
Note
: Do not use pointed objects for cleaning.
2 Manipulate moving parts.
Manipulate all moving parts such as the triggers, sleeves
and switches under running tap water to loosen and
remove gross debris.
Manual Cleaning Instructions
Mechanical/Automated Cleaning Instruction with Manual Pre-cleaning
Important:
– Manual pre-cleaning prior to mechanical/automated
cleaning/disinfection is important to ensure that
cannulations and other difficult to access areas are
clean.
– Alternative cleaning/disinfection procedures other
than in the procedure described below (including
manual precleaning) have not been validated by
Synthes.
3 Spray and wipe.
Spray and wipe the device using a neutral pH enzymatic
solution for a minimum of 2 minutes.
Follow the enzymatic detergent manufacturer’s
directions for correct temperature, water quality
(i.e. pH, hardness) and concentration/dilution.
4 Clean with detergent.
Clean the device manually under running warm water
using an enzymatic cleaner or detergent for a minimum
of 5 minutes. Manipulate all moving parts under running
water. Use a soft-bristled brush and/or soft lint-free cloth
to remove all visible soil and debris.
Follow the enzymatic cleaner or detergent
manufacturer’s instructions for use for correct
temperature, water quality and concentration/dilution.
5 Rinse with tap water.
Rinse the device thoroughly using cool to lukewarm
running water for a minimum of 2 minutes. Use a
syringe, pipette or water jet to flush lumens and
channels. Actuate joints, handles and other movable
device features in order to rinse thoroughly under
running water.
6 Visually inspect device.
Inspect the cannulations, coupling sleeves, etc. for visible
soil. Repeat steps 1 – 6 until no visible soil remains.
7
Final rinse with de-ionized/
purified water.
Final rinse with de-ionized or purified water for a
minimum of 2 minutes.
8 Dry.
Dry device using a soft lint-free cloth or clean
compressed air.
1 Remove debris.
Rinse the device under running cold tap water for a
minimum of 2 minutes. Use a sponge, soft lint-free
cloth or soft-bristled brush to assist in removing gross
soil. For cannulations of the handpiece and attachments,
the cleaning brush (519.40) should be used.
Note:
Do not use pointed objects for cleaning.
2 Manipulate moving parts.
Manipulate all moving parts such as the triggers, sleeves
and switches under running tap water to loosen and
remove gross debris.
3 Spray and wipe.
Spray and wipe the device using a neutral pH enzymatic
solution for a minimum of 2 minutes. Follow the
enzymatic detergent manufacturer’s directions for
correct temperature, water quality (i.e. pH, hardness)
and concentration/dilution.
4 Clean with detergent.
Clean the device manually under running warm water
using an enzymatic cleaner or detergent for a minimum
of 5 minutes. Manipulate all moving parts under running
water. Use a soft-bristled brush and/or soft lint-free cloth
to remove all visible soil and debris.
Follow the enzymatic cleaner or detergent manu-
facturer’s instructions for use for correct temperature,
water quality and concentration/dilution.
5 Rinse with tap water.
Rinse the device thoroughly using cool to lukewarm
running water for a minimum of 2 minutes. Use a
syringe, pipette or water jet to flush lumens and
channels. Actuate joints, handles and other movable
device features in order to rinse thoroughly under
running water.
6 Visually inspect device.
Inspect the cannulations, coupling sleeves, etc. for visible
soil. Repeat steps 1 – 6 until no visible soil remains.
7 Load washing basket.
Place devices in the specially designed tray for machine
washing supplied by Synthes (68.001.610). Ensure
that all cannulations (handpiece and attachments), if
applicable, are positioned vertically, i.e. in an upright
position as shown.
Note:
Specific lid (68.001.602) is available for the
washing basket.